Stahl A, Frick A, Imler M, Schlienger J L
Enzyme. 1979;24(5):294-301. doi: 10.1159/000458678.
Plasma L-alanine is deaminated by bacterial alanine-dehydrogenase; the resulting ammonia is dialyzed out and measured by use of a continuous flow phenol-hypochlorite colorimetric microassay. Concentrations in the range of 10-1,000 mumol alanine/l can be determined in a 50-microliter sample. The optimal conditions for the assay are specified. Study of the analytical qualities of the technique shows high specificity, good reproducibility , and a detection limit of 6 mumol/l. Usual values in human plasma from arterial or venous blood are respectively 296 +/- 166 and 376 +/- 214 mumol alanine/l (x +/- 2 SD). The usual values in rats are close to those found in man.
血浆L-丙氨酸由细菌丙氨酸脱氢酶脱氨;产生的氨经透析后,采用连续流动苯酚-次氯酸盐比色微量分析法进行测定。50微升样品中可测定10 - 1000微摩尔丙氨酸/升范围内的浓度。明确了该测定的最佳条件。对该技术分析质量的研究表明,其具有高特异性、良好的重现性,检测限为6微摩尔/升。动脉血或静脉血来源的人体血浆中通常值分别为296±166和376±214微摩尔丙氨酸/升(x±2标准差)。大鼠的通常值与人类的相近。