BACKGROUND

Few studies have been conducted regarding the optimal staple direction in gastrointestinal anastomosis. The purpose of this study was to evaluate the burst pressure of the anastomosis depending on the firing direction of the stapler.

METHODS

Pig esophagus and small bowel were used for all experiments. The small intestine represented a thin intestinal tract and the esophagus represented a thick intestinal tract. A side-to-side anastomosis was performed using a linear stapler, and the burst pressure was measured. A leak test on the anastomosis was then performed and burst pressures measured.

RESULTS

Burst pressures after anastomosis using a GIA™ 100-3.8 were 47.4 ± 10.4 mm Hg. With the same GIA, the burst pressure was significantly greater when the staples were driven from the small intestine into the esophagus (83.3 ± 17.3 mm Hg). Using the GIA™ 100-4.8, it was found that the burst pressure was significantly greater when the staplers were driven into the small intestine versus the esophagus (51.6 ± 7.1 vs. 68.6 ± 16.1 mm Hg). There was no significant difference between the different GIAs when fired in the same direction.

CONCLUSION

Burst pressures were significantly greater when the staplers were driven from the small intestine into the esophagus. The direction of the staple line influences the strength of the anastomosis.

摘要

背景

关于胃肠道吻合术中最佳吻合钉方向的研究较少。本研究的目的是根据吻合器的击发方向评估吻合口的爆破压力。

方法

所有实验均使用猪食管和小肠。小肠代表薄的肠道,食管代表厚的肠道。使用线性吻合器进行侧侧吻合,并测量爆破压力。然后对吻合口进行渗漏测试并测量爆破压力。

结果

使用GIA™100-3.8进行吻合后的爆破压力为47.4±10.4毫米汞柱。使用相同的GIA,当吻合钉从小肠打入食管时,爆破压力明显更高(83.3±17.3毫米汞柱)。使用GIA™100-4.8时,发现吻合器打入小肠时的爆破压力明显高于打入食管时(51.6±7.1对68.6±16.1毫米汞柱)。当以相同方向击发时,不同的GIA之间没有显著差异。

结论

当吻合器从小肠打入食管时,爆破压力明显更高。吻合钉线的方向会影响吻合口的强度。