Khoury A E, Olson M E, Lam K, Nickel J C, Costerton J W
Department of Biological Sciences, University of Calgary, Canada.
Br J Urol. 1989 Apr;63(4):384-8. doi: 10.1111/j.1464-410x.1989.tb05223.x.
We compared a newly developed closed urinary drainage system incorporating a retrograde contamination guard (RCG) with the conventional closed drainage system. The new system contains a solid bactericide (povidone iodine) pellet enclosed in a porous cartridge at the drain port of the urine collection bag. A catheterised rabbit model was used. The urine drainage bags were challenged daily for 8 days with an auxotrophically marked uropathogenic strain of Escherichia coli at the outlet tube. The bag urine was infected with the marked organism in 10 of the 11 rabbits in the control group (conventional collection bags) in 3.8 +/- 1.03 days and in only 1 of the 13 rabbits fitted with the RCG. The RCG effectively prevents contamination of the urine drainage bag and thus may play an important role in reducing the incidence of nosocomial urinary tract infections.
我们将一种新开发的带有逆行污染防护装置(RCG)的封闭式尿液引流系统与传统的封闭式引流系统进行了比较。新系统在尿液收集袋的排液口处有一个多孔药筒,药筒内装有固体杀菌剂(聚维酮碘)颗粒。采用了导尿兔模型。在8天的时间里,每天用营养缺陷型标记的致病性大肠杆菌菌株对尿液引流袋的出口管进行挑战。对照组(传统收集袋)的11只兔子中有10只在3.8±1.03天内其袋内尿液被标记的生物体感染,而在安装了RCG的13只兔子中只有1只被感染。RCG能有效防止尿液引流袋受到污染,因此可能在降低医院获得性尿路感染的发生率方面发挥重要作用。