[关于角膜前表面和后表面屈光力的Scheimpflug断层扫描测量的可重复性]

[Reproducibility of Scheimpflug Tomography Measurements Regarding Corneal Front and Back Surface Power].

作者信息

Stavridis E, Eppig T, Szentmáry N, Seitz B, Langenbucher A

机构信息

Klinik für Augenheilkunde, Universitätsklinikum des Saarlandes, Homburg/Saar.

Institut für Experimentelle Ophthalmologie, Universität des Saarlandes, Homburg/Saar.

出版信息

Klin Monbl Augenheilkd. 2015 Nov;232(11):1297-303. doi: 10.1055/s-0035-1558092. Epub 2015 Nov 17.

Abstract

BACKGROUND AND PURPOSE

The Oculus Pentacam® is one of the most commonly used devices in ophthalmology for assessment of the anterior eye segment. The purpose of this study was to determine the reproducibility of the corneal power as measured by the Pentacam® in a normal population.

METHODS

We enrolled 25 eyes of 25 subjects aged between 13 and 68 (46.7 ± 21.7) years, within a spherical equivalent from - 4 D to + 4 D and refractive cylinder up to 5 D and without notable pathologies or history of surgery. A sequence of 5 measurements was performed using the Pentacam®, the Zeiss IOLMaster® as well as an autorefractometer after (re-)positioning the patient's head and the measurement device. From the Pentacam® we collected power data for the corneal front and back surfaces, apical corneal thickness at apex and internal anterior chamber depth. From the IOLMaster®, we extracted corneal power data, and from the autorefractometer we obtained refractive power data of both cardinal meridians. For statistical analysis, we used Cronbach's α as a measure of reproducibility and Spearman's rank correlation test for correlation of data.

RESULTS

Pentacam® yields highly reproducible power data for both corneal surfaces, with Cronbach's α of greater than 0.97 for primary parameters (e.g. radii of curvature) and 0.9 for secondary parameters (e.g. vector components of astigmatism). Central corneal thickness and eccentricities of both surfaces as well as anterior chamber depth yielded highly reproducible values, with α of ≥ 0.97. Corneal power derived from IOLMaster® and objective refraction extracted from the autorefractometer yielded α values of around 0.9. Mean corneal power of Pentacam® and IOLMaster® correlated well, but there was no correlation to spherical equivalent. Astigmatism values from Pentacam® and IOLMaster® correlated well with each other and with refractive cylinder.

CONCLUSION

Our results indicate that measurements of anterior segment using the Pentacam® yield highly reproducible results in a normal population without major pathologies of the anterior eye segment.

摘要

背景与目的

Oculus Pentacam®是眼科最常用的用于评估眼前节的设备之一。本研究的目的是确定Pentacam®在正常人群中测量角膜屈光力的可重复性。

方法

我们纳入了25名年龄在13至68(46.7±21.7)岁之间的受试者的25只眼睛,等效球镜度在-4D至+4D之间,柱镜度高达5D,且无明显病变或手术史。在重新放置患者头部和测量设备后,使用Pentacam®、蔡司IOLMaster®以及自动验光仪进行了5次测量。从Pentacam®收集角膜前表面和后表面的屈光力数据、角膜顶点的角膜厚度以及前房深度。从IOLMaster®提取角膜屈光力数据,从自动验光仪获得两个主要子午线的屈光力数据。为了进行统计分析,我们使用Cronbach's α作为可重复性的指标,并使用Spearman秩相关检验来分析数据的相关性。

结果

Pentacam®对两个角膜表面的屈光力数据具有高度可重复性,主要参数(如曲率半径)的Cronbach's α大于0.97,次要参数(如散光的矢量分量)的Cronbach's α为0.9。中央角膜厚度、两个表面的偏心率以及前房深度产生了高度可重复的值,α≥0.97。IOLMaster®得出的角膜屈光力和自动验光仪得出的客观验光结果的α值约为0.9。Pentacam®和IOLMaster®的平均角膜屈光力相关性良好,但与等效球镜度无相关性。Pentacam®和IOLMaster®的散光值相互之间以及与柱镜度相关性良好。

结论

我们的结果表明,在没有眼前节重大病变的正常人群中,使用Pentacam®测量眼前节可产生高度可重复的结果。