J Cataract Refract Surg. 2016 Apr;42(4):569-76. doi: 10.1016/j.jcrs.2015.11.046.
To analyze the repeatability of corneal astigmatism measurements with 2 autokeratometers (IOLMaster 500 and Lenstar LS 900), 2 Placido topographers (Keratron Scout and Atlas 9000), and a Scheimpflug tomographer (Pentacam HR).
Department of Ophthalmology, Goethe University, Frankfurt, Germany.
Prospective case series.
The inclusion criteria were a minimum age of 18 years, corneal astigmatism of 3.00 diopters or less, and patient's consent to participate in the study. The main exclusion criteria were corneal pathologies, trauma, previous surgery, hard contact lens wear in the previous 3 months, and soft contact lens wear in the previous 6 weeks. In addition to manifest refraction and visual acuity measurements, all subjects had 2 full measurements of 1 eye using each device. Anterior corneal astigmatism and total corneal refractive power via calculations derived from the Scheimpflug tomographer were evaluated. Statistical analysis was performed using the Bland-Altman method and multiple linear regression analysis.
Forty-five eyes of 45 subjects (mean age 53.11 years ± 10.43 [SD]; 23 left eyes; 33 women) were included. The mean intermeasurement differences were approximately zero for all devices. The relative coefficient of repeatability of anterior corneal astigmatism, steep axis, and the vector analysis parameter J180 were lowest with the Scheimpflug tomographer. Total corneal refractive power measured with the Scheimpflug tomographer showed the highest repeatability in the larger diameter zones.
The Scheimpflug tomographer yielded higher repeatability of corneal astigmatism measurements, with total corneal refractive power measurements being as repeatable as established metrics.
Dr. Kohnen receives research support from Abbott Medical Optics Inc., Alcon Pharma GmbH, Bausch & Lomb GmbH, Carl Zeiss Meditec Vertriebsgesellschaft mbH, Hoya Surgical Optics GmbH, Oculus Optikgeräte GmbH, and Schwind eye-tech-solutions GmbH and Co. KG. He receives travel reimbursement, lecture fees and is a consultant to Abbott Medical Optics Inc., Alcon Pharma GmbH, Carl Zeiss Meditec Vertriebsgesellschaft mbH, Geuder AG, Oculus Optikgeräte GmbH, Schwind eye-tech-solutions GmbH and Co. KG, TearLab Corp., Thieme Compliance, and Ziemer Ophthalmic Systems AG. None of the other authors has a financial or proprietary interest in any of the materials or methods mentioned.
分析 2 台自动角膜曲率计(IOLMaster 500 和 Lenstar LS 900)、2 台角膜地形图仪(Keratron Scout 和 Atlas 9000)和 1 台 Scheimpflug 断层扫描仪(Pentacam HR)测量角膜散光的重复性。
德国法兰克福歌德大学眼科。
前瞻性病例系列。
纳入标准为年龄至少 18 岁,角膜散光 3.00 屈光度或以下,且患者同意参与研究。主要排除标准为角膜病变、创伤、既往手术、过去 3 个月内佩戴硬性隐形眼镜和过去 6 周内佩戴软性隐形眼镜。除了主觉验光和视力测量外,所有受试者均使用每种设备对 1 只眼进行 2 次完整测量。评估通过 Scheimpflug 断层扫描仪计算得出的前角膜散光和总角膜屈光力。采用 Bland-Altman 法和多元线性回归分析进行统计学分析。
纳入 45 名受试者(45 只眼)的数据分析,平均年龄 53.11 岁±10.43(SD);左眼 23 只;女性 33 名。所有设备的平均重复测量差值均接近零。Scheimpflug 断层扫描仪测量的前角膜散光、陡峭轴和向量分析参数 J180 的重复性最低。Scheimpflug 断层扫描仪测量的总角膜屈光力在较大直径区域的重复性最高。
Scheimpflug 断层扫描仪测量角膜散光的重复性更高,总角膜屈光力的测量与已建立的指标一样具有重复性。
Dr. Kohnen 从 Abbott Medical Optics Inc.、Alcon Pharma GmbH、Bausch & Lomb GmbH、Carl Zeiss Meditec Vertriebsgesellschaft mbH、Hoya Surgical Optics GmbH、Oculus Optikgeräte GmbH、Schwind eye-tech-solutions GmbH and Co. KG 获得研究支持。他还获得了旅行报销、演讲费,并担任 Abbott Medical Optics Inc.、Alcon Pharma GmbH、Carl Zeiss Meditec Vertriebsgesellschaft mbH、Geuder AG、Oculus Optikgeräte GmbH、Schwind eye-tech-solutions GmbH and Co. KG、TearLab Corp.、Thieme Compliance 和 Ziemer Ophthalmic Systems AG 的顾问。其他作者均与文中提到的任何材料或方法无财务或所有权利益。