Gao Y C, Jiang Y F, Lin S, Tian F
Eye Institute and School of Optometry, Tianjin Medical University Eye Hospital, Tianjin Key Laboratory of Retinal Functions and Diseases, Tianjin International Joint Research and Development Centre of Ophthalmology and Vision Science, Tianjin 300384, China.
Zhonghua Yan Ke Za Zhi. 2021 Jan 11;57(1):48-55. doi: 10.3760/cma.j.cn112142-20200904-00574.
To compare the difference and agreement of corneal refractive power and astigmatism measured by the new swept-source anterior segment optical coherence tomography (SS-AS-OCT) device (CASIA2) and Scheimpflug imaging device (Pentacam) in age-related cataract patients. In this cross-sectional study, 112 eyes of 112 patients with age-related cataract were examined before phacoemulsification in the Tianjin Medical University Eye Hospital from April to May 2020. The steep keratometry (Ks), flat keratometry (Kf), mean keratometry (Km), degree and axis of astigmatism of the corneal anterior and posterior surfaces and the total cornea were recorded. The difference in astigmatism was analyzed by the arithmetic method and the vector method. The difference of data was evaluated using Paired test or Wilcoxon test. The agreement of data was evaluated using Bland-Altman plots. The patients were 44 males and 68 females with an average age of (67±10) years. There were no statistically differences in the Ks and Km values of the corneal anterior surface between the Pentacam and the CASIA2 (both >0.05). There was significant difference in the Kf values of the corneal anterior surface between the Pentacam and the CASIA2 [(44.24±1.73) D . (44.14±1.64) D; =2.278; <0.05]. The Ks, Kf, and Km values of the posterior surface of the cornea measured by the Pentacam and the CASIA2 were (-6.60±0.29) D . (-6.45±0.28) D, (-6.34±0.27) D . (-6.17±0.25) D, and (-6.47±0.26) D . (-6.31±0.25) D, respectively. The Ks, Kf, and Km values of the total cornea measured by the two instruments were (45.08±1.84) D . (43.94±1.64) D, (44.18±1.85) D . (43.02±1.64) D, and (44.63±1.82) D . (43.48±1.60) D, respectively. There were significant differences in the Ks, Kf, and Km values of the posterior surface of the cornea and the total cornea (=-14.440, -13.522, -17.186, 21.016, 21.819, 22.981; all <0.01). The degree and axis of astigmatism of the corneal anterior and posterior surfaces and the total cornea showed no statistically significant difference (all >0.05). Vector operation results showed that the astigmatism difference vector (DV) of the Pentacam and the CASIA2 on the anterior surface of the cornea was 0.06 D@57°±0.64 D,>0.50 D in 47 eyes (41.96%). The astigmatism DV on the posterior surface of the cornea was 0.07 D@174°±0.21 D, >0.50 D in 2 eyes (1.79%). The astigmatism DV on the total cornea was 0.13 D@3°±0.69 D,>0.50 D in 59 eyes (52.68%). The results of the two devices were positively correlated ( values of the Ks, Kf, Km, and astigmatism degree on the anterior surface of the cornea were 0.970, 0.968, 0.976, and 0.697, respectively, on the posterior surface of the cornea were 0.918, 0.875, 0.925, and 0.517, respectively, and on the total cornea were 0.951, 0.955, 0.959, and 0.622, respectively; all <0.01). Bland-Altman analysis showed that the Ks, Kf, Km, and astigmatism degree of the corneal anterior and posterior surfaces measured by the two devices were consistent, with 2.68% to 8.04% of the measured values outside 95% limits of agreement (95%LoA). The 95%LoA of the total corneal Ks, Kf, and Km between the two devices was 0.01 to 2.28 D, 0.06 to 2.27 D, and 0.10 to 2.20 D, respectively, with a wide 95%LoA range and poor consistency. The corneal refractive power and astigmatism measured by the CASIA2 and the Pentacam shows little difference on the anterior surface of the cornea, with good agreement. However, the refractive power results of the posterior surface of the cornea and the total cornea show great difference, suggesting that these two instruments cannot be used interchangeably in clinical practice. .
比较新型扫频源眼前节光学相干断层扫描(SS - AS - OCT)设备(CASIA2)和Scheimpflug成像设备(Pentacam)在年龄相关性白内障患者中测量的角膜屈光力和散光的差异及一致性。在这项横断面研究中,于2020年4月至5月在天津医科大学眼科医院对112例年龄相关性白内障患者的112只眼进行了超声乳化术前检查。记录角膜前表面和后表面以及全角膜的陡峭角膜曲率(Ks)、平坦角膜曲率(Kf)、平均角膜曲率(Km)、散光度数和轴位。采用算术法和矢量法分析散光差异。使用配对t检验或Wilcoxon检验评估数据差异。使用Bland - Altman图评估数据一致性。患者中男性44例,女性68例,平均年龄(67±10)岁。Pentacam和CASIA2测量的角膜前表面Ks和Km值无统计学差异(均>0.05)。Pentacam和CASIA2测量的角膜前表面Kf值有显著差异[(44.24±1.73)D 与(44.14±1.64)D;t = 2.278;P < 0.05]。Pentacam和CASIA2测量的角膜后表面Ks、Kf和Km值分别为(-6.60±0.29)D 与(-6.45±0.28)D、(-6.34±0.27)D 与(-6.17±0.25)D、(-6.47±0.26)D 与(-6.31±0.25)D。两种仪器测量的全角膜Ks、Kf和Km值分别为(45.08±1.84)D 与(43.94±1.64)D、(44.18±1.85)D 与(43.02±1.64)D、(44.63±1.82)D 与(43.48±1.60)D。角膜后表面和全角膜的Ks、Kf和Km值有显著差异(t = -14.440,-13.522,-17.186,P = 21.016,21.819,22.981;均P < 0.01)。角膜前表面和后表面以及全角膜的散光度数和轴位无统计学显著差异(均>0.05)。矢量运算结果显示,Pentacam和CASIA2在角膜前表面的散光差异矢量(DV)为0.06 D@57°±0.64 D,47只眼(41.96%)>0.50 D。角膜后表面的散光DV为0.07 D@174°±0.21 D,2只眼(1.79%)>0.50 D。全角膜的散光DV为0.13 D@3°±0.69 D,59只眼(52.68%)>0.50 D。两种设备的结果呈正相关(角膜前表面Ks、Kf、Km和散光度数的r值分别为0.970、0.968、0.976和0.697,角膜后表面分别为0.918、0.875、0.925和0.517,全角膜分别为0.951、0.955、0.959和0.622;均P < 0.01)。Bland - Altman分析显示,两种设备测量的角膜前表面和后表面的Ks、Kf、Km和散光度数一致,2.68%至8.04%的测量值超出95%一致性界限(95%LoA)。两种设备全角膜Ks、Kf和Km的95%LoA分别为0.01至2.28 D、0.06至2.27 D和0.10至2.20 D,9 LoA范围较宽且一致性较差。CASIA2和Pentacam测量的角膜屈光力和散光在角膜前表面差异较小,一致性良好。然而,角膜后表面和全角膜的屈光力结果差异较大,提示这两种仪器在临床实践中不能互换使用。