Suppr超能文献

家庭健康中使用的无液血压计的设备校准率。

Aneroid Blood Pressure Manometer Calibration Rates of Devices Used in Home Health.

作者信息

Arena Sara K, Bacyinski Andrew, Simon Lee, Peterson Edward L

机构信息

Sara K. Arena, PT, DScPT, is an Assistant Professor, School of Health Sciences, Oakland University, Rochester, Michigan; and Physical Therapist, Henry Ford Health System, Community Care Services, Detroit, Michigan. Andrew Bacyinski, SPT, is a Physical Therapy Student, School of Health Sciences, Oakland University, Rochester, Michigan. Lee Simon, SPT, is a Physical Therapy Student, School of Health Sciences, Oakland University, Rochester, Michigan. Edward L. Peterson, PhD, is a Senior Biostatistician, Department of Public Health, Henry Ford Health System, Detroit, Michigan.

出版信息

Home Healthc Now. 2016 Jan;34(1):23-8. doi: 10.1097/NHH.0000000000000321.

Abstract

Hypertension is associated with cardiovascular disease, stroke, and a range of other medical sequelae. Accurate blood pressure (BP) readings, which depend on the integrity and calibration of the measuring device, are essential to identifying suboptimal BP. This study describes calibration rates of aneroid BP devices (a) utilized in home healthcare (HHC) and (b) having the needle resting within the zero accuracy indicator. BP devices from one branch of a home care agency were inspected and checked for calibration according to the protocol set forth by the European Society of Hypertension. Of the 125 devices measured, 78.4% were in calibration. Of the 94 devices with the gauge needle resting in the zero accuracy indicator, 11.7% were not in calibration; whereas, 51.6% of the 31 devices with the gauge needle resting outside the zero accuracy indicator were found not in calibration. Twenty-one devices were not checked for calibration due to inflation bulb malfunction, tubing tears, or excessive wear. Furthermore, visual inspection of the needle placement did not confirm a device as being in or out of calibration. Proper maintenance and routine calibration of BP equipment is foundational to assuring accuracy of BP readings obtained by HHC providers.

摘要

高血压与心血管疾病、中风以及一系列其他医学后遗症相关。准确的血压(BP)读数依赖于测量设备的完整性和校准,对于识别未达最佳水平的血压至关重要。本研究描述了(a)家庭医疗保健(HHC)中使用的、(b)指针位于零精度指示器内的无液血压计的校准率。根据欧洲高血压学会制定的方案,对一家家庭护理机构一个分支机构的血压计进行了检查和校准检查。在测量的125台设备中,78.4%处于校准状态。在94台指针位于零精度指示器内的设备中,11.7%未校准;而在31台指针位于零精度指示器外的设备中,51.6%未校准。由于充气球囊故障、管道撕裂或过度磨损,有21台设备未进行校准检查。此外,通过目视检查指针位置并不能确定设备是否校准。血压设备的正确维护和定期校准是确保家庭医疗保健提供者获得准确血压读数的基础。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验