基层医疗中临床使用的血压计压力刻度的准确性。

Accuracy of the pressure scale of sphygmomanometers in clinical use within primary care.

作者信息

Coleman Andrew J, Steel Stephen D, Ashworth Mark, Vowler Sarah L, Shennan Andrew

机构信息

Medical Physics Department, Division of Reproductive Health, Endocrinology and Development, Kings College School of Medicine, Guy's & St Thomas' NHS Foundation Trust, London SE1 7EH cGKT, UK.

出版信息

Blood Press Monit. 2005 Aug;10(4):181-8. doi: 10.1097/01.mbp.0000168398.87167.c2.

Abstract

BACKGROUND

It is widely recommended that sphygmomanometers are maintained and calibrated regularly to ensure that the pressure scale remains accurate to within the European Standard specification of +/-3 mmHg. In primary care, however, such checks are reported to be only rarely performed. This paper describes a survey of the accuracy of the absolute static pressure scale of aneroid, mercury and automated sphygmomanometers in clinical use in primary care.

METHODS

On-site measurements of sphygmomanometer pressure scale accuracy were carried out in 45 general practices within Lambeth, Southwark and Lewisham. A total of 279 sphygmomanometers from these practices were included in the study. The device pressure scales were calibrated using an accurate electronic reference pressure sensor.

RESULTS

The key finding of this study is that 17.9% (50 out of 279) of all surveyed devices gave errors exceeding the +/-3 mmHg threshold. Of these, 53.2% (33 out of 62) of aneroid devices were found to be reading in error by more than +/-3 mmHg compared with 7.8% (16 out of 217) of the combined population of mercury and automated devices. The difference between these groups is statistically significant (P=0.002). Significant differences in the performance of specific models of aneroid, mercury and automated devices were also identified.

CONCLUSION

A service model for improving the accuracy of blood pressure monitoring in primary care needs to take into account the current proliferation of pressure scale errors in these devices, the lack of uptake of regular checks and the poor quality of some of the devices currently in use.

摘要

背景

广泛建议定期维护和校准血压计,以确保压力刻度在欧洲标准规范的±3 mmHg范围内保持准确。然而,在初级医疗保健中,据报道此类检查很少进行。本文描述了一项对初级医疗保健临床使用的无液、汞柱式和自动血压计绝对静压刻度准确性的调查。

方法

在兰贝斯、南华克和刘易舍姆的45家全科诊所对血压计压力刻度准确性进行现场测量。这些诊所共有279台血压计纳入研究。使用精确的电子参考压力传感器对设备压力刻度进行校准。

结果

本研究的主要发现是,所有被调查设备中有17.9%(279台中的50台)误差超过±3 mmHg阈值。其中,发现53.2%(62台中的33台)无液设备读数误差超过±3 mmHg,而汞柱式和自动设备的综合群体中这一比例为7.8%(217台中的16台)。这些组之间的差异具有统计学意义(P = 0.002)。还发现了特定型号的无液、汞柱式和自动设备在性能上的显著差异。

结论

改善初级医疗保健中血压监测准确性的服务模式需要考虑到这些设备目前压力刻度误差的普遍存在、定期检查的缺乏以及目前使用的一些设备质量较差的情况。