使用血糖仪比较昏迷中毒患者的毛细血管血糖水平和静脉血糖水平以及实验室血糖水平。

Comparison of capillary and venous blood glucose levels using glucometer and laboratory blood glucose level in poisoned patients being in coma.

作者信息

Yaraghi Ahmad, Mood Nastaran Eizadi, Dolatabadi Leila Kamali

机构信息

Department of Anesthesiology, Isfahan University of Medical Science, Isfahan, Iran.

Isfahan Clinical Toxicology Research Center, Isfahan University of Medical Science, Isfahan, Iran.

出版信息

Adv Biomed Res. 2015 Nov 23;4:247. doi: 10.4103/2277-9175.170242. eCollection 2015.

Abstract

BACKGROUND

Poisoning is one of the most common medical presentations in a hospital. Hypoglycemic patients are at increased risk of toxicity. The purpose of this study was to compare capillary blood glucose and venous blood glucose measurements using glucometer against laboratory blood glucose in case of poisoned patients being in coma.

MATERIALS AND METHODS

During the 6-month study period, a random sample of 98 patients was admitted in the Department of Poisoning Emergency and Clinical Toxicology of Noor Hospital, Isfahan University of Medical Sciences, Iran from May 2010. Data collected included age, gender, poisoning reason, vital signs, and Glasgow Coma Scale. Capillary blood samples were obtained from the fourth fingertip of the non-dominant hand. t-Test, paired t-test, Pearson's correlation analysis, and one-way analysis of variance (ANOVA) were used for analysis.

RESULTS

The mean of capillary blood glucose was 115.7 ± 50.2, of venous blood glucose measured by glucose meter was 117.8 ± 47.3, and of glucose measured in vitro was 115.8 ± 55.1. Mean of blood glucose showed no significant difference with the three mentioned methods. The correlation between capillary and intravenous blood glucose samples measured by glucometer was 0.93, between capillary blood glucose and in vitro measured venous blood glucose was 0.78, and between venous blood glucose measured by glucose meter and in vitro measured sample was 0.81. The mean of capillary and venous blood glucose levels measured by glucose meter, capillary and venous blood glucose levels measured in vitro, and venous blood glucose levels measured by glucose meter had no significant differences.

CONCLUSION

Using venous blood sample and measuring the glucose level in it by glucometer is an acceptable and advisable method, and capillary blood glucose measurement by using glucometer is not recommended for patients in coma.

摘要

背景

中毒是医院中最常见的医疗情况之一。低血糖患者中毒风险增加。本研究的目的是比较在中毒昏迷患者中使用血糖仪测量的毛细血管血糖和静脉血糖与实验室血糖测量结果。

材料与方法

在为期6个月的研究期间,从2010年5月起,对伊朗伊斯法罕医科大学努尔医院中毒急诊与临床毒理学科收治的98例患者进行随机抽样。收集的数据包括年龄、性别、中毒原因、生命体征和格拉斯哥昏迷量表。从非优势手的无名指指尖采集毛细血管血样。采用t检验、配对t检验、Pearson相关分析和单因素方差分析(ANOVA)进行分析。

结果

毛细血管血糖平均值为115.7±50.2,血糖仪测量的静脉血糖平均值为117.8±47.3,体外测量的血糖平均值为115.8±55.1。三种测量方法的血糖平均值无显著差异。血糖仪测量的毛细血管血糖与静脉血糖样本之间的相关性为0.93,毛细血管血糖与体外测量的静脉血糖之间的相关性为0.78,血糖仪测量的静脉血糖与体外测量样本之间的相关性为0.81。血糖仪测量的毛细血管和静脉血糖水平、体外测量的毛细血管和静脉血糖水平以及血糖仪测量的静脉血糖水平之间无显著差异。

结论

使用静脉血样本并通过血糖仪测量其中的血糖水平是一种可接受且可取的方法,不建议对昏迷患者使用血糖仪测量毛细血管血糖。