A new non-linear approach was applied to calculate the left ventricular ejection fraction (LVEF) using multigated acquisition (MUGA) images.
In this study, 50 patients originally for the estimation of the percentage of LVEF to monitor the effects of various cardiotoxic drugs in chemotherapy were retrospectively selected. All patients had both MUGA and echocardiography examinations (ECHO LVEF) at the same time. Mutual information (MI) theory was utilized to calculate the LVEF using MUGA imaging (MUGA MI).
MUGA MI estimation was significantly different from MUGA LVEF and ECHO LVEF, respectively (p < 0.005). The higher repeatability for MUGA MI can be observed in the figure by the higher correlation coefficient for MUGA MI (r = 0.95) compared with that of MUGA LVEF (r = 0.80). Again, the reproducibility was better for MUGA MI (r = 0.90, 0.92) than MUGA LVEF (r = 0.77, 0.83). The higher correlation coefficients were obtained between proposed MUGA MI and ECHO LVEF compared to that between the conventional MUGA LVEF and ECHO LVEF.
MUGA image with the aid of MI is promising to be more interchangeable LVEF to ECHO LVEF measurement as compared with the conventional approach on MUGA image.