Masoud Ahmed I, Bulic Milena, Viana Grace, Bedran-Russo Ana K
a Faculty member, Department of Orthodontics, Faculty of Dentistry, King Abdulaziz University, Jeddah, Saudi Arabia; and Clinical Assistant Professor, Department of Orthodontics, College of Dentistry, University of Illinois at Chicago, Chicago, IL.
b Orthodontic Resident, Department of Orthodontics, College of Dentistry, University of Illinois at Chicago, Chicago, IL.
Angle Orthod. 2016 Sep;86(5):818-25. doi: 10.2319/082815-581.1. Epub 2016 Jan 7.
To compare over a period of 8 weeks (1) the force decay and (2) the dimensional changes between thermoplastic (TP) and thermoset (TS) elastomeric ligatures.
TP and TS elastomeric ligatures were obtained from Rocky Mountain Orthodontics™. The TS ligatures were custom made specifically for this study. The sample included 72 clear TP and 72 clear TS elastomeric ligatures. The experiment was performed in a simulated oral environment (pH of 6.75) at 37°C. The remaining forces and the dimensional changes were measured at different time points over a period of 8 weeks.
Student's t-tests revealed significant differences in percent force loss, percent change in outer diameter, percent change in inner diameter, and percent change in wall thickness between TP and TS elastomeric ligatures across all time points (P < .001). The difference in percent change in width between TP and TS elastomeric ligatures was not significant at all time points (P > .05). The mean difference in force loss between TP and TS across all time points was 22.91%. The TP and TS specimens exhibited 93.04% and 77.41% force loss, respectively, at the 28th day.
This novel TS elastomeric ligature showed significantly less force decay and dimensional changes over time; therefore, it might be superior during initial leveling and aligning and during finishing stages. Using a transfer jig to prevent relaxation of the specimens before force measurement showed that force decay of commercially available elastomeric ligatures was greater than that described in previous publications.
在8周的时间内比较(1)热塑性(TP)和热固性(TS)弹性结扎丝之间的力衰减情况,以及(2)尺寸变化情况。
TP和TS弹性结扎丝购自落基山正畸公司(Rocky Mountain Orthodontics™)。TS结扎丝是专门为本研究定制的。样本包括72根透明TP弹性结扎丝和72根透明TS弹性结扎丝。实验在37°C、pH值为6.75的模拟口腔环境中进行。在8周的时间内,于不同时间点测量剩余力和尺寸变化。
学生t检验显示,在所有时间点上,TP和TS弹性结扎丝在力损失百分比、外径变化百分比、内径变化百分比和壁厚变化百分比方面存在显著差异(P < .001)。TP和TS弹性结扎丝在宽度变化百分比上的差异在所有时间点均不显著(P > .05)。所有时间点上TP和TS之间的平均力损失差异为22.91%。在第28天时,TP和TS样本的力损失分别为93.04%和77.41%。
这种新型TS弹性结扎丝随时间推移显示出显著更少的力衰减和尺寸变化;因此,在初始整平排齐阶段和精细调整阶段可能更具优势。在力测量前使用转移夹具防止样本松弛,结果显示市售弹性结扎丝的力衰减大于先前出版物中所述。