Bone attachment to hydroxyapatite coated polymers.

作者信息

Boone P S, Zimmerman M C, Gutteling E, Lee C K, Parsons J R, Langrana N

机构信息

Department of Orthopedic Surgery, University of Medicine and Dentistry of New Jersey, Newark 07103.

出版信息

J Biomed Mater Res. 1989 Aug;23(A2 Suppl):183-99.

PMID:2674147
Abstract

Bone attachment to two classes of hydroxyapatite (HA) coated polymers was evaluated mechanically and histologically. Particulate HA was molded into the surface of thermoplastic implants and cast into the surface of thermoset implants. Coated and uncoated implants of both types were implanted in the distal femurs of rabbits for four and twelve weeks. After sacrifice, the bone/implant interface was evaluated mechanically via a push-out test with a servohydraulic test system. Paired statistical analysis revealed significantly greater shear strengths for coated vs. uncoated implants for both polymer systems at both time periods. Mixed mode of failure occurred with particle removal from the polymer surface and from the bone. One animal at each time period was saved for histology. Histology at both time periods indicated direct bone apposition to the HA coating as compared to a fibrous encapsulation about the uncoated implants. This was confirmed with backscatter scanning electron microscopy and scanning acoustic microscopy. Hydroxyapatite coatings significantly improve the bond strength between polymers and bone by allowing direct bone apposition and some mechanical interlocking with the bone.

摘要

对两类羟基磷灰石(HA)涂层聚合物的骨附着情况进行了力学和组织学评估。将颗粒状HA模塑到热塑性植入物表面,并浇铸到热固性植入物表面。两种类型的涂层和未涂层植入物均植入兔股骨远端,为期4周和12周。处死后,通过伺服液压测试系统的推出试验对骨/植入物界面进行力学评估。配对统计分析显示,在两个时间段内,两种聚合物系统的涂层植入物与未涂层植入物相比,剪切强度均显著更高。失效模式为混合模式,表现为颗粒从聚合物表面和骨中脱出。每个时间段各留一只动物用于组织学检查。两个时间段的组织学检查均表明,与未涂层植入物周围的纤维包裹相比,骨直接附着于HA涂层。背散射扫描电子显微镜和扫描声学显微镜证实了这一点。羟基磷灰石涂层通过允许骨直接附着并与骨进行一些机械互锁,显著提高了聚合物与骨之间的结合强度。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验