Truant A L, Starr E, Nevel C A, Tsolakis M, Fiss E F
Department of Microbiology and Immunology, Temple University School of Medicine, Philadelphia, PA.
Diagn Microbiol Infect Dis. 1989 May-Jun;12(3):211-5. doi: 10.1016/0732-8893(89)90017-5.
Identification of Gram-negative bacilli by AMS-Vitek, MicroScan, and Autobac Series II was evaluated with 434 clinical isolates comprising glucose fermenters and glucose nonfermenters. The organisms were tested in each of the three systems (the AMS Vitek GNI card, the MicroScan Combo Plus Panel, and the Autobac 18 chamber cuvette). API 20E was used as the primary reference system. MicroScan and Vitek correctly identified 96.1% and 95.6% of the organisms, respectively. The Autobac Series II identified 82.3% of the organisms correctly. Results with MicroScan, Vitek, and Autobac were available in 24-48 hours, 4-18 hours, and 3-8 hours, respectively.
采用包含葡萄糖发酵菌和葡萄糖非发酵菌的434株临床分离株,对AMS-Vitek、MicroScan和Autobac II系列进行革兰氏阴性杆菌鉴定评估。这些菌株在三个系统(AMS Vitek GNI卡、MicroScan Combo Plus Panel和Autobac 18孔比色皿)中分别进行测试。API 20E用作主要参考系统。MicroScan和Vitek分别正确鉴定了96.1%和95.6%的菌株。Autobac II系列正确鉴定了82.3%的菌株。MicroScan、Vitek和Autobac的结果分别在24-48小时、4-18小时和3-8小时内可得。