Park Kwangwoo, Choi Wonhoon, Keum Ki Chang, Lee Ho, Yoon Jeongmin, Lee Chang Geol, Lee Ik Jae, Cho Jaeho
Department of Radiation Oncology, College of Medicine, Yonsei University, Seoul, Republic of Korea.
Department of Radiation Oncology, College of Medicine, Yonsei University, Seoul, Republic of Korea.
Phys Med. 2016 Feb;32(2):398-402. doi: 10.1016/j.ejmp.2016.01.003. Epub 2016 Jan 28.
Mechanical Quality Assurance (QA) is important to assure spatially precise delivery of external-beam radiation therapy. As an alternative to the conventional-film based method, we have developed a new tool for mechanical QA of LINACs which uses a light field rather than radiation. When light passes through the collimator, a shadow is projected onto a piece of translucent paper and the resulting image is captured by a digital camera via a mirror. With this method, we evaluated the position of the LINAC isocenter and the accuracy of the gantry, collimator, and couch rotation. We also evaluated the accuracy of the digital readouts of the gantry, collimator, and couch rotation. In addition, the treatment couch position indicator was tested. We performed camera calibration as an essential pre-requisite for quantitative measurements of the position of isocenter, the linear motion of the couch, and the rotation angles of the gantry and collimator. Camera calibration reduced the measurement error to submillimeter based on uncertainty in pixel size of the image, while, without calibration, the measurement error of up to 2 mm could occur for an object with a length of 5 cm.
机械质量保证(QA)对于确保外照射放射治疗在空间上的精确输送非常重要。作为基于传统胶片方法的替代方案,我们开发了一种用于直线加速器机械QA的新工具,该工具使用光场而非辐射。当光线穿过准直器时,会在一张半透明纸上投射出一个阴影,所得图像通过镜子由数码相机捕获。通过这种方法,我们评估了直线加速器等中心的位置以及机架、准直器和治疗床旋转的精度。我们还评估了机架、准直器和治疗床旋转数字读数的准确性。此外,还对治疗床位置指示器进行了测试。我们进行了相机校准,这是对等中心位置、治疗床直线运动以及机架和准直器旋转角度进行定量测量的必要前提。基于图像像素大小的不确定性,相机校准将测量误差降低到了亚毫米级别,而未进行校准的情况下,对于长度为5厘米的物体,测量误差可能高达2毫米。