González A, Castro I, Martínez J A
Servicio de Radiofísica y Protección Radiológica, Hospital Universitario Virgen de la Arrixaca, Ctra. Madrid-Cartagena S/N, El Palmar (Murcia) E-30120, Spain.
Med Phys. 2004 Jun;31(6):1489-93. doi: 10.1118/1.1755491.
Measurement of radiation isocenter is a fundamental part of commissioning and quality assurance (QA) for a linear accelerator (linac). In this work we present an automated procedure for the analysis of the stars-shots employed in the radiation isocenter determination. Once the star-shot film has been developed and digitized, the resulting image is analyzed by scanning concentric circles centered around the intersection of the lasers that had been previously marked on the film. The center and the radius of the minimum circle intersecting the central rays are determined with an accuracy and precision better than 1% of the pixel size. The procedure is applied to the position and size determination of the radiation isocenter by means of the analysis of star-shots, placed in different planes with respect to the gantry, couch and collimator rotation axes.
辐射等中心的测量是直线加速器(linac)调试和质量保证(QA)的基本组成部分。在这项工作中,我们提出了一种自动程序,用于分析辐射等中心确定中使用的星射影像。一旦星射影像胶片显影并数字化,通过扫描以先前在胶片上标记的激光束交点为中心的同心圆来分析所得图像。与中心射线相交的最小圆的中心和半径的确定精度优于像素大小的1%。通过分析相对于机架、治疗床和准直器旋转轴放置在不同平面的星射影像,该程序应用于辐射等中心的位置和大小确定。