波动掩蔽器对高性能人工耳蜗使用者言语理解的影响。

The effect of fluctuating maskers on speech understanding of high-performing cochlear implant users.

作者信息

Zirn Stefan, Polterauer Daniel, Keller Stefanie, Hemmert Werner

机构信息

a University of Applied Science , Offenburg , Germany .

b Department of Otorhinolaryngology of the Medical Center , University of Freiburg , Germany .

出版信息

Int J Audiol. 2016;55(5):295-304. doi: 10.3109/14992027.2015.1128124. Epub 2016 Feb 10.

Abstract

OBJECTIVE

The present study evaluated whether the poorer baseline performance of cochlear implant (CI) users or the technical and/or physiological properties of CI stimulation are responsible for the absence of masking release.

DESIGN

This study measured speech reception thresholds (SRTs) in continuous and modulated noise as a function of signal to noise ratio (SNR).

STUDY SAMPLE

A total of 24 subjects participated: 12 normal-hearing (NH) listeners and 12 subjects provided with recent MED-EL CI systems.

RESULTS

The mean SRT of CI users in continuous noise was -3.0 ± 1.5 dB SNR (mean ± SEM), while the normal-hearing group reached -5.9 ± 0.8 dB SNR. In modulated noise, the difference across groups increased considerably. For CI users, the mean SRT worsened to -1.4 ± 2.3 dB SNR, while it improved for normal-hearing listeners to -18.9 ± 3.8 dB SNR.

CONCLUSIONS

The detrimental effect of fluctuating maskers on SRTs in CI users shown by prior studies was confirmed by the current study. Concluding, the absence of masking release is mainly caused by the technical and/or physiological properties of CI stimulation, not just the poorer baseline performance of many CI users compared to normal-hearing subjects. Speech understanding in modulated noise was more robust in CI users who had a relatively large electrical dynamic range.

摘要

目的

本研究评估了人工耳蜗(CI)使用者较差的基线表现或CI刺激的技术和/或生理特性是否是导致掩蔽释放缺失的原因。

设计

本研究测量了在连续噪声和调制噪声中作为信噪比(SNR)函数的言语接受阈值(SRT)。

研究样本

共有24名受试者参与:12名听力正常(NH)的听众和12名配备了最新MED-EL CI系统的受试者。

结果

CI使用者在连续噪声中的平均SRT为-3.0±1.5 dB SNR(平均值±标准误),而听力正常组达到-5.9±0.8 dB SNR。在调制噪声中,组间差异显著增加。对于CI使用者,平均SRT恶化至-1.4±2.3 dB SNR,而听力正常的听众则改善至-18.9±3.8 dB SNR。

结论

本研究证实了先前研究中波动掩蔽器对CI使用者SRT的有害影响。结论是,掩蔽释放的缺失主要是由CI刺激的技术和/或生理特性引起的,而不仅仅是许多CI使用者与听力正常受试者相比较差的基线表现。在具有相对较大电动态范围的CI使用者中,调制噪声中的言语理解更强。