Jeal D E
Br J Anaesth. 1977 Mar;49(3):273-6. doi: 10.1093/bja/49.3.273.
A simple method of adapting a pressure-cycled ventilator of the flow generator type to ventilate patients on a circle-absorber anaesthetic system is described. Standard lengths of corrugated black rubber tubing are placed between the ventilator and the circle-absorber is deadspace, replacing the reservoir bag. The performance of this modification was investigated and a functional volume for the tubing calculated. There was a simple relationship between the tidal volume, the length of tubing and the fresh gas flow required to avoid dilution of the anaesthetic gases by the driving gases.
描述了一种将流量发生器型压力循环呼吸机适配于在循环吸收式麻醉系统上对患者进行通气的简单方法。在呼吸机和循环吸收器之间放置标准长度的波纹状黑色橡胶管以形成无效腔,取代贮气囊。对这种改进的性能进行了研究,并计算了该管道的功能容积。潮气量、管道长度与避免麻醉气体被驱动气体稀释所需的新鲜气体流量之间存在简单的关系。