Teijeira F J, Lamoureux G, Tétreault J P, Bauset R, Guidoin R, Marois Y, Paynter R, Assayed F
Chirurgie Cardiothoracique et Médecine Nucléaire, CHUS, Sherbrooke, Québec, Canada.
Biomaterials. 1989 Mar;10(2):80-4. doi: 10.1016/0142-9612(89)90035-5.
Hydrophilic, microporous polyurethane vascular prostheses displayed good mechanical characteristics and the behaviour in vitro was excellent. The in-vivo results were, however, disappointing. To obtain a better understanding of the phenomena involved in the acute and early thrombosis, we implanted the polyurethane graft in the canine left femoral artery and an autologous femoral vein in the right for 4 and 24 h. At 4 h, one polyurethane graft was totally occluded and the other two were close to complete occlusion; at 24 h, none was patent. On the other hand, all autologous veins were patent. The thrombotic matrix incorporating both platelets and fibrinogen, quantified by labelling, was anchored along the anastomotic lines. This study confirms that polyurethane occlusion is initiated by hyperplastic reaction, but does not explain why.
亲水性微孔聚氨酯血管移植物表现出良好的机械特性,体外性能也很出色。然而,体内实验结果却令人失望。为了更好地理解急性和早期血栓形成所涉及的现象,我们将聚氨酯移植物植入犬的左股动脉,并将自体股静脉植入右侧,分别放置4小时和24小时。4小时时,一个聚氨酯移植物完全闭塞,另外两个接近完全闭塞;24小时时,所有移植物均无通畅。另一方面,所有自体静脉均保持通畅。通过标记定量的包含血小板和纤维蛋白原的血栓基质,沿吻合线固定。本研究证实聚氨酯闭塞是由增生反应引发的,但并未解释其原因。