在CT血管造影期间,采用非侵入性心输出量测量作为试验团注技术的替代方法。

A non-invasive cardiac output measurement as an alternative to the test bolus technique during CT angiography.

作者信息

Nijhof W H, Hilbink M, Jager G J, Slump C H, Rutten M J C M

机构信息

University of Twente, MIRA-Institute for Biomedical Technology and Technical Medicine, Drienerlolaan 5, 7522 NB, Enschede, The Netherlands; Department of Radiology, Jeroen Bosch Hospital, Henri Dunantstraat 1, 5223 GZ, 's-Hertogenbosch, The Netherlands.

Jeroen Bosch Hospital, Henri Dunantstraat 1, 5223 GZ, 's-Hertogenbosch, The Netherlands.

出版信息

Clin Radiol. 2016 Sep;71(9):940.e1-5. doi: 10.1016/j.crad.2016.03.007. Epub 2016 Apr 10.

Abstract

AIM

To investigate the association between a non-invasive cardiac output (CO) measurement and the scan delay, as derived from a test bolus injection protocol. The secondary objective was to determine which factors affect the relationship between the CO and scan delay.

MATERIALS AND METHODS

Fifty-five patients referred for a contrast-enhanced (thorax-)abdomen CT examination were included in this feasibility study. A test bolus examination was performed prior to the abdominal CT. During the test bolus injection, the CO of the patient was measured using a non-invasive finger-cuff measurement. Associations were analysed using linear regression analyses. Age, gender, height, weight, and blood pressure were included as potential confounders.

RESULTS

Linear regression analysis showed a negative and significant association between CO and delay. The regression formula was as follows: scan delay (seconds) = 26.8-1.6 CO (l/min), with a 95% CI between -2.3 and -1.0 (p<0.001). Weight appeared to be a confounder in this relation, and gender and blood pressure were effect modifiers. There was no interaction between scan delay and age, height and weight.

CONCLUSIONS

There is a negative and significant association between the non-invasive CO measurement and the CT scan delay; however, to validate these findings a larger cohort study is needed to investigate whether the non-invasively determined scan delay is as accurate as the use of a test bolus.

摘要

目的

研究通过试验性团注注射方案得出的非侵入性心输出量(CO)测量值与扫描延迟之间的关联。次要目的是确定哪些因素会影响CO与扫描延迟之间的关系。

材料与方法

本可行性研究纳入了55名因胸部-腹部增强CT检查而转诊的患者。在腹部CT检查前进行了试验性团注检查。在试验性团注注射期间,使用非侵入性手指袖带测量法测量患者的CO。使用线性回归分析来分析关联。将年龄、性别、身高、体重和血压作为潜在的混杂因素纳入分析。

结果

线性回归分析显示CO与延迟之间存在显著的负相关。回归公式如下:扫描延迟(秒)=26.8 - 1.6×CO(升/分钟),95%置信区间在-2.3至-1.0之间(p<0.001)。体重似乎是这种关系中的一个混杂因素,而性别和血压是效应修饰因素。扫描延迟与年龄、身高和体重之间没有交互作用。

结论

非侵入性CO测量值与CT扫描延迟之间存在显著的负相关;然而,为了验证这些发现,需要进行更大规模的队列研究,以调查非侵入性确定的扫描延迟是否与使用试验性团注一样准确。