Suppr超能文献

一项在日本进行的观察性研究,比较原型设备与现有设备在老年患者中有效可视化不可见静脉的效果。

An observational study comparing the prototype device with the existing device for the effective visualization of invisible veins in elderly patients in Japan.

作者信息

Kimori Keiko, Sugama Junko, Nakatani Toshio, Nakayama Kazuya, Miyati Tosiaki, Sanada Hiromi

机构信息

Ishikawa Prefectural Nursing University, Kahoku, Japan.

Wellness Promotion Science Center, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Kanazawa, Japan.

出版信息

SAGE Open Med. 2015 Nov 27;3:2050312115615365. doi: 10.1177/2050312115615365. eCollection 2015.

Abstract

OBJECTIVE

To compare the performance on the detection of the invisible veins between our modified prototype device and an existing device in elderly hospitalized patients.

METHODS

A prospective, cross-sectional, and observational study was performed in the invisible veins in elderly patients. The major variables, skin color near the invisible veins, and diameter and depth of the invisible veins were measured. The vein visualization rate was calculated as the ratio of the visualized veins to the invisible veins by the visualization device.

RESULTS

We analyzed 53 invisible veins in the cubital fossa and 56 invisible veins in the forearm in a total of 72 patients (median age, 73 years). The visualization rate for our prototype device was higher than that for an existing device in the cubital fossa and the forearm sites. The visualized veins of the prototype device had a higher intensity ratio than that of an existing device. No significant differences were observed in the body mass index, vein depth, and vein diameter of the visualized veins at the cubital fossa and forearm sites.

CONCLUSION

The prototype surpassed the existing device in visualizing the invisible veins. However, the prototype was unable to visualize all the invisible veins. We need to look for ways to reduce noise and to visualize the invisible veins, and the visualization rate of devices needs to be investigated in further association with the percentage of success with actual intravenous access and locating time to vein.

摘要

目的

比较改良后的原型设备与现有设备在老年住院患者中检测不可见静脉的性能。

方法

对老年患者的不可见静脉进行前瞻性、横断面观察研究。测量不可见静脉附近的皮肤颜色、不可见静脉的直径和深度等主要变量。静脉可视化率通过可视化设备将可视化静脉数与不可见静脉数之比计算得出。

结果

共纳入72例患者(中位年龄73岁),分析了肘窝处53条不可见静脉和前臂处56条不可见静脉。在肘窝和前臂部位,原型设备的可视化率高于现有设备。原型设备可视化静脉的强度比高于现有设备。在肘窝和前臂部位可视化静脉的体重指数、静脉深度和静脉直径方面未观察到显著差异。

结论

在可视化不可见静脉方面,原型设备优于现有设备。然而,原型设备无法可视化所有不可见静脉。我们需要寻找减少噪声和可视化不可见静脉的方法,并且需要进一步结合实际静脉穿刺成功百分比和静脉定位时间来研究设备的可视化率。