Hidalgo Irene Ruiz, Rozema Jos J, Dhubhghaill Sorcha Ní, Zakaria Nadia, Koppen Carina, Tassignon Marie-José
J Refract Surg. 2015 Mar;31(3):176-81. doi: 10.3928/1081597X-20150224-01.
To determine the repeatability of a color LED corneal topographer (Cassini; iOptics, The Hague, The Netherlands) and compare it with Placido and Scheimpflug based devices (EyeSys 2000; EyeSys Laboratories, Houston, TX, and Pentacam HR; Oculus Optikgeräte GmbH, Wetzlar, Germany).
This prospective study involved 20 healthy volunteers (20 eyes) recruited from the staff of the Antwerp University Hospital. For each eye, three measurements were taken using each device, from which eight parameters describing keratometry and astigmatism were derived. Repeatability was assessed using intrasession, within-subject analysis of variance. Agreement was evaluated between pairs of devices with Bland- Altman plots and 95% confidence intervals.
The repeatability values were good for all three devices, although slightly worse for the Cassini than for the other two devices for steep, flat, and mean keratometry. The EyeSys showed worse repeatability for the astigmatism axis. EyeSys and Pentacam agreed well with each other in terms of mean keratometry, whereas the Cassini gave consistently higher values by 0.52 diopters (D) when compared with the EyeSys (P < .05, paired t test) and by 0.38 D when compared with the Pentacam (P < .05, paired t test). The Cassini provided similar repeatability values (2.31°) for the astigmatism axis to the Pentacam (2.22°), in contrast to the EyeSys, which produced much more variable axes (9.0°) (P < .05, paired t test).
Overall, the three devices display comparable repeatability. The one exception is the astigmatism axis measurement of the EyeSys. The Cassini provides higher keratometry values than the other two devices, but the astigmatism axis agrees well with that provided by the Pentacam.
确定彩色发光二极管角膜地形图仪(卡西尼;iOptics,荷兰海牙)的可重复性,并将其与基于普拉西多和谢姆弗勒原理的设备(EyeSys 2000;EyeSys实验室,德克萨斯州休斯顿,以及Pentacam HR;德国韦茨拉尔的Oculus Optikgeräte GmbH)进行比较。
这项前瞻性研究纳入了从安特卫普大学医院工作人员中招募的20名健康志愿者(20只眼)。对于每只眼睛,使用每种设备进行三次测量,从中得出八个描述角膜曲率和散光的参数。使用测量期间内受试者方差分析评估可重复性。通过布兰德-奥特曼图和95%置信区间评估成对设备之间的一致性。
所有三种设备的可重复性值都很好,尽管卡西尼在陡峭、平坦和平均角膜曲率方面的可重复性略逊于其他两种设备。EyeSys在散光轴方面的可重复性较差。EyeSys和Pentacam在平均角膜曲率方面彼此一致性良好,而卡西尼与EyeSys相比,平均角膜曲率值始终高0.52屈光度(D)(P <.05,配对t检验),与Pentacam相比高0.38 D(P <.05,配对t检验)。卡西尼提供的散光轴可重复性值(2.31°)与Pentacam(2.22°)相似,而EyeSys产生的轴变化大得多(9.0°)(P <.05,配对t检验)。
总体而言,这三种设备显示出相当的可重复性。唯一的例外是EyeSys的散光轴测量。卡西尼提供的角膜曲率值高于其他两种设备,但散光轴与Pentacam提供的结果一致性良好。