Schug Titus, Kohnen Thomas, Kaiser Klemens Paul, Lwowski Christoph
Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.
Acta Ophthalmol. 2025 Aug;103(5):e310-e317. doi: 10.1111/aos.17487. Epub 2025 Mar 28.
To evaluate the effect of artificial tear (AT) application on the repeatability of corneal parameters with three different devices.
In this prospective, randomized clinical trial, 145 patients were measured with one of three devices in the Department of Ophthalmology, Goethe University, Frankfurt, Germany. Two baseline measurements and four measurements following AT application after 30 s, 2, 5 and 10 min were performed.
Forty-eight eyes were measured with the IOLMaster 700 (Carl Zeiss Meditec AG, Oberkochen, Germany) and Galilei G6 (Ziemer, Port, Switzerland), and 49 with the Pentacam AXL (Oculus, Wetzlar, Germany) (145 total). Repeatability of the mean corneal power (K) between baseline measurements was high (0.049 ± 0.008 D for IOLMaster, 0.061 ± 0.011 for the Pentacam AXL and 0.065 ± 0.009 for Galilei G6) but significantly variable after 30 s and 2 min for IOLMaster 700 (0.177 ± 0.021 D, p < 0.001 and 0.079 ± 0.009, p = 0.004) and Galilei G6 (0.163 ± 0.020 D, p < 0.001 and 0.119 ± 0.014 D, p = 0.003). K differed significantly from baseline at 30 s, 2 and 5 min with Pentacam AXL (0.215 ± 0.041 D, p < 0.001; 0.157 ± 0.025 D, p < 0.001 and 0.094 ± 0.012 D, p = 0.007). Variability was highest at 30 s and decreased over time.
The use of AT leads to high variability of keratometric measurements for each device. This should be considered after use, and at least 5 min should elapse before measurements.
评估使用人工泪液(AT)对三种不同设备测量角膜参数重复性的影响。
在这项前瞻性随机临床试验中,德国法兰克福歌德大学眼科使用三种设备之一对145例患者进行测量。进行了两次基线测量以及在使用AT后30秒、2分钟、5分钟和10分钟时的四次测量。
使用IOLMaster 700(德国奥伯科亨卡尔蔡司医疗技术股份公司)和Galilei G6(瑞士波特齐默尔公司)测量了48只眼,使用Pentacam AXL(德国韦茨拉尔奥culus公司)测量了49只眼(共145只眼)。基线测量之间平均角膜屈光力(K)的重复性较高(IOLMaster为0.049±0.008 D,Pentacam AXL为0.061±0.011 D,Galilei G6为0.065±0.009 D),但IOLMaster 700在30秒和2分钟后变化显著(0.177±0.021 D,p<0.001;0.079±0.009,p = 0.004),Galilei G6也如此(0.163±0.020 D,p<0.001;0.119±0.014 D,p = 0.003)。Pentacam AXL在30秒、2分钟和5分钟时K与基线有显著差异(0.215±0.041 D,p<0.001;0.157±0.025 D,p<0.001;0.094±0.012 D,p = 0.007)。变异性在30秒时最高,并随时间降低。
使用AT会导致每种设备角膜测量的变异性较高。使用后应考虑到这一点,测量前至少应经过5分钟。