Ning Jiliang, Zhang Lijun
Department of Ophthalmology, The Third People's Hospital of Dalian, Dalian, China.
Department of Ophthalmology, Dalian Municipal Eye Hospital, Dalian, China.
Front Med (Lausanne). 2024 Nov 5;11:1490674. doi: 10.3389/fmed.2024.1490674. eCollection 2024.
This study aimed to compare the consistency of pupil offset measurements obtained using the Pentacam, Keratron Scout, and Sirius devices.
This retrospective cross-sectional study included 146 young myopic individuals (292 eyes) scheduled for refractive surgery at Dalian Third People's Hospital between January 2023 and December 2023. Three devices were utilized to measure the chord mu of the pupil deviation along with the Cartesian distances of the X and Y coordinates (Px, Py) associated with the pupil offset. Repeated-measures analysis of variance was used to compare differences in pupil offset acquisition across various devices. Additionally, the intraclass correlation coefficient (ICC) and Bland-Altman plot were utilized to assess the consistency among the three devices.
Chord mu, measured using the Pentacam, Keratron Scout, and Sirius devices, were 0.18 ± 0.10, 0.21 ± 0.11, and 0.18 ± 0.11, respectively. The Px values were 0.00 ± 0.14, -0.02 ± 0.16, and -0.01 ± 0.13, respectively, while the Py values were 0.09 ± 0.13, 0.10 ± 0.15, and 0.10 ± 0.13. The ICCs for the three device measurements, chord mu, Px, and Py, were 0.817, 0.900, and 0.855, respectively. When comparing the three devices, the 95% limits of agreement (LoA) for mu and Px measured using the Sirius and Keratron Scout were the narrowest, ranging from -0.15 to 0.08 and -0.11 to 0.13, respectively. Additionally, the 95% LoA for Py measured using the Sirius and Pentacam was the narrowest, ranging from -0.13 to 0.15. The pupil centers in both eyes were predominantly located above the apex of the cornea.
Sirius, Keratron Scout, and Pentacam have good consistency in pupil shift measurement in young myopic patients, and the three devices can be used as references in clinical practice.
本研究旨在比较使用Pentacam、Keratron Scout和Sirius设备获得的瞳孔偏移测量结果的一致性。
这项回顾性横断面研究纳入了2023年1月至2023年12月期间在大连第三人民医院计划进行屈光手术的146名年轻近视患者(292只眼)。使用三种设备测量瞳孔偏差的弦长μ以及与瞳孔偏移相关的X和Y坐标(Px,Py)的笛卡尔距离。采用重复测量方差分析来比较不同设备在瞳孔偏移测量上的差异。此外,使用组内相关系数(ICC)和Bland-Altman图来评估这三种设备之间的一致性。
使用Pentacam、Keratron Scout和Sirius设备测量的弦长μ分别为0.18±0.10、0.21±0.11和0.18±0.11。Px值分别为0.00±0.14、-0.02±0.16和-0.01±0.13,而Py值分别为0.09±0.13、0.10±0.15和0.10±0.13。三种设备测量的弦长μ、Px和Py的ICC分别为0.817、0.900和0.855。在比较这三种设备时,使用Sirius和Keratron Scout测量的μ和Px的95%一致性界限(LoA)最窄,分别为-0.15至0.08和-0.11至0.13。此外,使用Sirius和Pentacam测量的Py的95% LoA最窄,为-0.13至0.15。双眼的瞳孔中心主要位于角膜顶点上方。
Sirius、Keratron Scout和Pentacam在年轻近视患者的瞳孔移位测量中具有良好的一致性,这三种设备可在临床实践中作为参考。