Grant Alexander M, Deller Timothy W, Khalighi Mohammad Mehdi, Maramraju Sri Harsha, Delso Gaspar, Levin Craig S
Department of Bioengineering, Stanford University, Stanford, California 94305-5128 and Department of Radiology, Stanford University, Stanford, California 94305-5128.
GE Healthcare, Waukesha, Wisconsin 53188-1678.
Med Phys. 2016 May;43(5):2334. doi: 10.1118/1.4945416.
The GE SIGNA PET/MR is a new whole body integrated time-of-flight (ToF)-PET/MR scanner from GE Healthcare. The system is capable of simultaneous PET and MR image acquisition with sub-400 ps coincidence time resolution. Simultaneous PET/MR holds great potential as a method of interrogating molecular, functional, and anatomical parameters in clinical disease in one study. Despite the complementary imaging capabilities of PET and MRI, their respective hardware tends to be incompatible due to mutual interference. In this work, the GE SIGNA PET/MR is evaluated in terms of PET performance and the potential effects of interference from MRI operation.
The NEMA NU 2-2012 protocol was followed to measure PET performance parameters including spatial resolution, noise equivalent count rate, sensitivity, accuracy, and image quality. Each of these tests was performed both with the MR subsystem idle and with continuous MR pulsing for the duration of the PET data acquisition. Most measurements were repeated at three separate test sites where the system is installed.
The scanner has achieved an average of 4.4, 4.1, and 5.3 mm full width at half maximum radial, tangential, and axial spatial resolutions, respectively, at 1 cm from the transaxial FOV center. The peak noise equivalent count rate (NECR) of 218 kcps and a scatter fraction of 43.6% are reached at an activity concentration of 17.8 kBq/ml. Sensitivity at the center position is 23.3 cps/kBq. The maximum relative slice count rate error below peak NECR was 3.3%, and the residual error from attenuation and scatter corrections was 3.6%. Continuous MR pulsing had either no effect or a minor effect on each measurement.
Performance measurements of the ToF-PET whole body GE SIGNA PET/MR system indicate that it is a promising new simultaneous imaging platform.
GE SIGNA PET/MR是通用电气医疗集团推出的一款新型全身一体化飞行时间(ToF)PET/MR扫描仪。该系统能够以低于400皮秒的符合时间分辨率同时进行PET和MR图像采集。同时进行PET/MR作为一种在一项研究中探究临床疾病分子、功能和解剖参数的方法具有巨大潜力。尽管PET和MRI具有互补的成像能力,但由于相互干扰,它们各自的硬件往往不兼容。在这项工作中,对GE SIGNA PET/MR的PET性能以及MRI运行干扰的潜在影响进行了评估。
遵循NEMA NU 2 - 2012协议测量PET性能参数,包括空间分辨率、噪声等效计数率、灵敏度、准确性和图像质量。这些测试中的每一项都是在MR子系统空闲以及在PET数据采集期间进行连续MR脉冲的情况下进行的。大多数测量在系统安装的三个不同测试地点重复进行。
在距横断面视野中心1厘米处,扫描仪在径向、切向和轴向的半高宽空间分辨率分别平均达到4.4毫米、4.1毫米和5.3毫米。在活度浓度为17.8 kBq/ml时,达到了218 kcps的峰值噪声等效计数率(NECR)和43.6%的散射分数。中心位置的灵敏度为23.3 cps/kBq。低于峰值NECR时的最大相对切片计数率误差为3.3%,衰减和散射校正的残余误差为3.6%。连续MR脉冲对每项测量要么没有影响,要么影响较小。
ToF - PET全身GE SIGNA PET/MR系统的性能测量表明它是一个有前景的新型同时成像平台。