Suppr超能文献

使用多层电离室的笔形束质子射线照相术。

Pencil beam proton radiography using a multilayer ionization chamber.

作者信息

Farace Paolo, Righetto Roberto, Meijers Arturs

机构信息

Proton therapy Unit, APSS, Hospital of Trento, Via del Desert, 14, Trento, Italy.

出版信息

Phys Med Biol. 2016 Jun 7;61(11):4078-87. doi: 10.1088/0031-9155/61/11/4078. Epub 2016 May 10.

Abstract

A pencil beam proton radiography (PR) method, using a commercial multilayer ionization chamber (MLIC) integrated with a treatment planning system (TPS) was developed. A Giraffe (IBA Dosimetry) MLIC (±0.5 mm accuracy) was used to obtain pencil beam PR by delivering spots uniformly positioned at a 5.0 mm distance in a 9  ×  9 square of spots. PRs of an electron-density (with tissue-equivalent inserts) phantom and a head phantom were acquired. The integral depth dose (IDD) curves of the delivered spots were computed by the TPS in a volume of water simulating the MLIC, and virtually added to the CT at the exit side of the phantoms. For each spot, measured and calculated IDD were overlapped in order to compute a map of range errors. On the head-phantom, the maximum dose from PR acquisition was estimated. Additionally, on the head phantom the impact on the range errors map was estimated in case of a 1 mm position misalignment. In the electron-density phantom, range errors were within 1 mm in the soft-tissue rods, but greater in the dense-rod. In the head-phantom the range errors were  -0.9  ±  2.7 mm on the whole map and within 1 mm in the brain area. On both phantoms greater errors were observed at inhomogeneity interfaces, due to sensitivity to small misalignment, and inaccurate TPS dose computation. The effect of the 1 mm misalignment was clearly visible on the range error map and produced an increased spread of range errors (-1.0  ±  3.8 mm on the whole map). The dose to the patient for such PR acquisitions would be acceptable as the maximum dose to the head phantom was  <2cGyE. By the described 2D method, allowing to discriminate misalignments, range verification can be performed in selected areas to implement an in vivo quality assurance program.

摘要

开发了一种铅笔束质子射线照相(PR)方法,该方法使用与治疗计划系统(TPS)集成的商用多层电离室(MLIC)。使用长颈鹿(IBA剂量学)MLIC(精度为±0.5毫米),通过在9×9的点正方形中以5.0毫米的距离均匀分布点来获得铅笔束PR。获取了电子密度(带有组织等效插入物)体模和头部体模的PR。通过TPS在模拟MLIC的水体积中计算所输送点的积分深度剂量(IDD)曲线,并在体模出口侧虚拟添加到CT上。对于每个点,将测量的和计算的IDD重叠,以计算射程误差图。在头部体模上,估计了PR采集的最大剂量。此外,在头部体模上估计了1毫米位置未对准情况下对射程误差图的影响。在电子密度体模中,软组织棒中的射程误差在1毫米以内,但在致密棒中更大。在头部体模中,整个图上的射程误差为-0.9±2.7毫米,在脑区内在1毫米以内。在两个体模上,由于对小的未对准敏感以及TPS剂量计算不准确,在不均匀性界面处观察到更大的误差。1毫米未对准的影响在射程误差图上清晰可见,并导致射程误差的分布增加(整个图上为-1.0±3.8毫米)。由于头部体模的最大剂量<2cGyE,因此这种PR采集对患者的剂量是可以接受的。通过所描述的二维方法,可以区分未对准情况,从而在选定区域进行射程验证,以实施体内质量保证计划。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验