Nelson J M, Huff J W, Bennett E S, Davis L J
University of Missouri School of Optometry, St. Louis.
CLAO J. 1989 Apr-Jun;15(2):125-33.
Variation in thickness is known to affect the Dk/L of rigid gas permeable contact lenses. Our study was designed to evaluate the variability of central and average thickness of Oxyflow EW lenses. In addition, the relationship between overnight swelling of the central cornea and Dk/L was examined to determine whether Dk/L calculated with central lens thickness correlated better with lens-induced edema than Dk/L calculated with average lens thickness. The results showed marked variation among lens center thicknesses for all powers (-6 D, -2 D, +2 D, +6 D) examined, and, as expected, more variation for centrally calculated Dk/L with high minus lenses. Average Dk/L did not vary significantly with power. Average and central Dk/L values had a near 1:1 relationship for -2 D lenses for the base curve (7.8 mm) ordered; but only when power was kept constant was a good correlation seen between central and average Dk/L. A one-patient study evaluating the overnight central swelling response to three +6 D and three -6 D (high, medium, and low center thickness) lenses showed that central Dk/L correlated better with lens-induced edema than average Dk/L. Thus, specifying central thickness on lens orders appears to be a primary variable influencing extended wear edema at night. Peripheral Dk/L, however, may influence corneal physiological variables other than central edema and warrants further study.
众所周知,厚度变化会影响硬性透气性角膜接触镜的Dk/L值。我们的研究旨在评估Oxyflow EW镜片中央厚度和平均厚度的变异性。此外,还研究了中央角膜过夜肿胀与Dk/L之间的关系,以确定用镜片中央厚度计算的Dk/L是否比用平均镜片厚度计算的Dk/L与镜片引起的水肿相关性更好。结果显示,在所检查的所有屈光度(-6D、-2D、+2D、+6D)的镜片中心厚度之间存在显著差异,并且,正如预期的那样,高负度数镜片中央计算的Dk/L差异更大。平均Dk/L不会随屈光度显著变化。对于所订购的基弧为7.8mm的-2D镜片,平均Dk/L值与中央Dk/L值几乎呈1:1的关系;但只有当屈光度保持恒定时,中央Dk/L与平均Dk/L之间才会呈现出良好的相关性。一项针对一名患者的研究评估了其对三片+6D和三片-6D(高、中、低中心厚度)镜片的过夜中央肿胀反应,结果表明,中央Dk/L比平均Dk/L与镜片引起的水肿相关性更好。因此,在镜片订单上指定中央厚度似乎是影响夜间长期佩戴水肿的一个主要变量。然而,周边Dk/L可能会影响除中央水肿之外的角膜生理变量,值得进一步研究。