Hashemi Hassan, Asgari Soheila, Mortazavi Mahdi, Ghaffari Reza
Noor Ophthalmology Research Center (H.H., S.A.), Noor Eye Hospital, Tehran, Iran; and Eye Research Center (M.M., R.G.), Farabi Eye Hospital, Tehran University of Medical Sciences, Tehran, Iran.
Eye Contact Lens. 2017 Nov;43(6):371-377. doi: 10.1097/ICL.0000000000000280.
To compare the effect of femtosecond-assisted thin flap laser-assisted in situ keratomileusis (FS-LASIK) and photorefractive keratectomy with mitomycin-C (PRK-MMC) in highly myopic patients (>7.0 D [D]) on corneal biomechanical parameters.
In this prospective comparative interventional case series, 60 patients (30 patients in each group) with a manifest refractive spherical equivalent (MRSE) greater than 7.0 D were enrolled. Corvis ST parameters were measured before and at 3 and 6 months postoperatively.
Preoperatively, mean MRSE was -8.65±1.51 D in FS-LASIK and -8.04±1.70 D in PRK-MMC groups (P=0.149), and corneal thickness was 570.67±36.79 μm and 507.12±32.55 μm, respectively (P<0.001). At 6 months, both groups showed significantly higher applanation (A) 2 time and A2-velocity (P<0.05 in both), whereas intraocular pressure (IOP), corrected IOP, A1 time, A1-velocity, and radius at highest concavity were significantly reduced (P<0.05 in all). In the FS-LASIK group, there was a significant increase in deformation amplitude (DA) (P=0.001), and significant decreases in A2-length (P=0.004). Peak distance increased in the PRK-MMC group (P=0.029). At 6 months, after controlling for fellow eye correlations and preoperative corneal thickness between the two groups, decreases in IOP, A1-time, A2-length and radius, and the increase in DA was greater in FS-LASIK.
This study demonstrated significant changes in Corvis ST ocular biomechanical metrics after both PRK-MMC and FS-LASIK in high myopic patients, indicating the significant effect of excimer laser refractive surgery on corneal biomechanical properties. However, changes that occur with FS-LASIK are more significant than with PRK-MMC. Further randomized studies are needed to better characterize the pattern of biomechanical changes associated with each type of surgery.
比较飞秒辅助薄瓣准分子原位角膜磨镶术(FS-LASIK)和丝裂霉素C准分子激光角膜切削术(PRK-MMC)对高度近视患者(>7.0 D [屈光度])角膜生物力学参数的影响。
在这个前瞻性比较干预性病例系列中,纳入了60例患者(每组30例),其明显屈光球镜等效度(MRSE)大于7.0 D。在术前以及术后3个月和6个月测量Corvis ST参数。
术前,FS-LASIK组的平均MRSE为-8.65±1.51 D,PRK-MMC组为-8.04±1.70 D(P = 0.149);角膜厚度分别为570.67±36.79μm和507.12±32.55μm(P < 0.001)。在6个月时,两组的压平(A)2次时间和A2速度均显著升高(两者P < 0.05),而眼压(IOP)、校正眼压、A1时间、A1速度和最高凹陷处半径均显著降低(所有P < 0.05)。在FS-LASIK组中,变形幅度(DA)显著增加(P = 0.00`1),A2长度显著降低(P = 0.004)。PRK-MMC组的峰值距离增加(P = 0.029)。在6个月时,在控制了两组之间的对侧眼相关性和术前角膜厚度后,FS-LASIK组眼压、A1时间、A2长度和半径的降低以及DA的增加幅度更大。
本研究表明,PRK-MMC和FS-LASIK术后高度近视患者的Corvis ST眼生物力学指标均有显著变化,表明准分子激光屈光手术对角膜生物力学特性有显著影响。然而,FS-LASIK引起的变化比PRK-MMC更显著。需要进一步的随机研究来更好地描述与每种手术相关的生物力学变化模式。