Zhang Ting, Chen Lin, Huang Jianpan, Li Jing, Cai Shuhui, Cai Congbo, Chen Zhong
Department of Electronic Science, Fujian Provincial Key Laboratory of Plasma and Magnetic Resonance, Xiamen University, Xiamen, China.
Department of Communication Engineering, Xiamen University, Xiamen, China.
J Magn Reson. 2016 Aug;269:138-145. doi: 10.1016/j.jmr.2016.06.002. Epub 2016 Jun 4.
As a recently emerging method, spatiotemporally encoded (SPEN) magnetic resonance imaging (MRI) has a high robustness to field inhomogeneity and chemical shift effect. It has been broadened from single-slice scanning to multi-slice scanning. In this paper, a novel multi-slice SPEN MRI method was proposed. In this method, the slice-selective dimension was segmented to lower the specific absorption rate (SAR) and improve the image quality. This segmented method, dubbed SeSPEN method, was theoretically analyzed and demonstrated with phantom, lemon and in vivo rat brain experiments. The experimental results were compared with the results obtained from the spin-echo EPI, spin-echo SPEN method and multi-slice global SPEN method proposed by Frydman and coauthors (abbr. GlSPEN method). All the SPEN images were super-resolved reconstructed using deconvolution method. The results indicate that the SeSPEN method retains the advantage of SPEN MRI with respect to resistance to field inhomogeneity and can provide better signal-to-noise ratio than multi-slice GlSPEN MRI technique. The SeSPEN method has comparable SAR to the GlSPEN method while the T1 signal attenuation effect is alleviated. The proposed method will facilitate the multi-slice SPEN MRI to scan more slices within one scan with better image quality.
作为一种新兴方法,时空编码(SPEN)磁共振成像(MRI)对场不均匀性和化学位移效应具有很高的鲁棒性。它已从单切片扫描扩展到多切片扫描。本文提出了一种新颖的多切片SPEN MRI方法。在该方法中,对切片选择维度进行分割以降低比吸收率(SAR)并提高图像质量。这种分割方法被称为SeSPEN方法,并通过体模、柠檬和活体大鼠脑实验进行了理论分析和验证。将实验结果与自旋回波EPI、自旋回波SPEN方法以及Frydman及其合作者提出的多切片全局SPEN方法(简称GlSPEN方法)所获得的结果进行了比较。所有SPEN图像均使用去卷积方法进行超分辨率重建。结果表明,SeSPEN方法保留了SPEN MRI在抵抗场不均匀性方面的优势,并且能够提供比多切片GlSPEN MRI技术更好的信噪比。SeSPEN方法的SAR与GlSPEN方法相当,同时减轻了T1信号衰减效应。所提出的方法将有助于多切片SPEN MRI在一次扫描中扫描更多切片并具有更好的图像质量。