Suppr超能文献

压力下的支气管阻塞器:体外模型和离体模型。

Bronchial blockers under pressure: in vitro model and ex vivo model.

机构信息

Unit of Anesthesia, Intensive Care and Pain Management, Department of Medicine

Unit of Anesthesia, Intensive Care and Pain Management, Department of Medicine.

出版信息

Br J Anaesth. 2016 Sep;117 Suppl 1:i92-i96. doi: 10.1093/bja/aew120. Epub 2016 Jun 14.

Abstract

BACKGROUND

Pressures (Pe) exerted by bronchial blockers on the inner wall of the bronchi may cause mucosal ischaemia. Our aims were as follows: (i) to compare the intracuff pressure (Pi) and Pe exerted by commercially available bronchial blockers in an in vitro and an ex vivo model; (ii) to investigate the influence of both the inflated intracuff volume and cuff diameter on Pe; and (iii) to estimate the minimal sealing volume (VSmin) and the corresponding Pe for each bronchial blocker studied.

METHODS

The Pe exerted by seven commercial bronchial blockers was measured at different inflation volumes using a custom-designed system using in vitro and ex vivo animal models with two internal diameters (12 and 15 mm).

RESULTS

In the same conditions, Pi was significantly lower than Pe (P<0.05), and Pe was higher in the in vitro model than in the ex vivo model. The Pe increased with the inflated volume, with use of the small-diameter model (P<0.05). Ex vivo models needed a higher minimal sealing volume than the in vitro models, and this volume increased with the diameter (e.g. the VSmin at a positive pressure of 25 cm H2O required a Pe ranging from 12 to 78 mm Hg on the 15 mm ex vivo model and from 66 to 110 mm Hg on the 12 mm ex vivo model).

CONCLUSIONS

The Pi cannot be used to approximate Pe. The diameter of the model, the inflated volume, and the bronchial blocker design all influence Pe. A pressure higher than the critical ischaemic threshold (i.e. 25 mm Hg) was needed to prevent air leak around the cuff in the in vitro and ex vivo models.

摘要

背景

支气管阻塞器对内壁的压力(Pe)可能导致黏膜缺血。我们的目的如下:(i)在体外和离体模型中比较市售支气管阻塞器的套囊内压(Pi)和 Pe;(ii)研究套囊充气量和直径对 Pe 的影响;(iii)估计每个研究的支气管阻塞器的最小密封容积(VSmin)和相应的 Pe。

方法

使用定制的系统,在体外和离体动物模型(内径分别为 12 和 15mm)中,使用不同的充气量测量了 7 种商业支气管阻塞器的 Pe。

结果

在相同条件下,Pi 显著低于 Pe(P<0.05),且体外模型中的 Pe 高于离体模型。Pe 随充气量的增加而增加,使用小直径模型时(P<0.05)。离体模型需要比体外模型更高的最小密封容积,并且该容积随直径的增加而增加(例如,在 15mm 离体模型上,正压为 25cmH2O 时,VSmin 需要的 Pe 范围为 12 至 78mmHg,在 12mm 离体模型上,Pe 范围为 66 至 110mmHg)。

结论

Pi 不能用来近似 Pe。模型的直径、充气量和支气管阻塞器的设计都会影响 Pe。为了防止在体外和离体模型中套囊周围发生空气泄漏,需要高于临界缺血阈值(即 25mmHg)的压力。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验