Minerva Anestesiol. 2017 Jan;83(1):23-32. doi: 10.23736/S0375-9393.16.11135-6. Epub 2016 Jun 17.
Gross morphological differences exist among different brands of pediatric supraglottic devices (SGDs). The aim of this study is to compare the spatial relationship of i-gel® and Ambu® AuraOnce (AO)TM on pediatric airway based on three dimensional (3-D) magnetic resonance imaging (MRI) measurements.
Sixty patients up to 12 years of age were enrolled and assigned in two groups, i-gel® or Ambu® AOTM. After confirmation of proper placement of these SGDs, 3-D MRI scans of head and neck were performed. Another native scan was also obtained after removal of the SGD for comparison.
i-gel® produced significant degree of compression of the tongue (P<0.001) while Ambu® AOTM significantly reduced the axial diameter of glottis (P=0.033) compared to their native values. Both i-gel® and Ambu® AOTM significantly reduced the area of the glottic opening (P<0.001 for each device) and the distance between the arytenoids (P<0.001 and P=0.007 respectively); and increased the distance between the hyoid bone and cervical spine (P<0.001 and P=0.001 respectively) in comparison to their corresponding native values. Bowl of i-gel® produced greater dilation of the upper esophageal sphincter at all levels of measurement- upper (P<0.001), middle (P=0.001) and lower (P=0.015) in comparison to Ambu® AOTM.
Based on 3-D MRI measurements done on living patients, both SGDs distorted the anatomy of pediatric airway compared to their respective native values to variable extent. The relevance of these effects needs further studies on larger patient group in order to reduce morbidity on pediatric airway.
不同品牌的小儿声门上装置(SGD)之间存在明显的宏观形态差异。本研究旨在通过三维(3-D)磁共振成像(MRI)测量比较 i-gel®和 Ambu® AuraOnce(AO)TM 在小儿气道中的空间关系。
共纳入 60 例年龄在 12 岁以下的患者,分为 i-gel®组和 Ambu®AOTM 组。确认这些 SGD 正确放置后,对头部和颈部进行 3-D MRI 扫描。在移除 SGD 后,还获得了另一个原始扫描以进行比较。
与原始值相比,i-gel®显著压缩了舌部(P<0.001),而 Ambu®AOTM 显著减小了声门的轴向直径(P=0.033)。与原始值相比,i-gel®和 Ambu®AOTM 均显著减小了声门开口面积(P<0.001,每个装置)和杓状软骨之间的距离(P<0.001 和 P=0.007);并且与相应的原始值相比,增加了舌骨和颈椎之间的距离(P<0.001 和 P=0.001)。与 Ambu®AOTM 相比,i-gel®的杯形部分在所有测量水平上都使上食管括约肌扩张更大-上部(P<0.001),中部(P=0.001)和下部(P=0.015)。
基于对活体患者进行的 3-D MRI 测量,与各自的原始值相比,两种 SGD 都不同程度地扭曲了小儿气道的解剖结构。为了减少小儿气道的发病率,需要在更大的患者群体中进一步研究这些影响的相关性。