Assailly J, Monet J D, Moustafa Y, Brugal G
INSERM, Hôpital Necker-Enfants Malades, Paris.
J Urol (Paris). 1989;95(1):3-9.
Image processors and present specific software applications have now placed scanning microcytometry in the ranks of other conventional technics for routine laboratory examinations, particularly for urinary cytology and DNA measurements (ploidy). Conventional urinary sediment smears of 52 patients with bladder tumor were digitized, processed and statistically analyzed by the image analyzer, Samba. They were automatically classified and ploidy measurements were made and compared with 6 conventional cytological diagnostics for each smear. The Samba program assigns the cell images to two classes, negative and positive. 100% of the cells were correctly classed with respect to the cytologist's diagnostics. Within the positive class the degree of malignancy established by the program was the same as the cytologist's classifications. For smears of class III, the comparison of the diagnostics of the cytologists for each patient permitted the validation of the differential classification, negative-positive, made using quantitative cytology. The interpretation of the two DNA histogram parameters, the degree of ploidy and the proliferation index, provided an excellent prognosis of which patients would show tumoral recrudescence, as verified by follow ups.
图像处理器及特定的软件应用程序现已将扫描微细胞计量术提升至常规实验室检查的其他传统技术行列,尤其适用于尿液细胞学检查和DNA测量(倍体分析)。使用图像分析仪Samba对52例膀胱肿瘤患者的常规尿沉渣涂片进行数字化处理、分析及统计。自动分类并进行倍体测量,并将每个涂片的结果与6种传统细胞学诊断结果进行比较。Samba程序将细胞图像分为阴性和阳性两类。就细胞学家的诊断而言,100%的细胞被正确分类。在阳性类别中,该程序确定的恶性程度与细胞学家的分类一致。对于III类涂片,对每位患者细胞学家的诊断结果进行比较,证实了使用定量细胞学进行的阴性-阳性鉴别分类。对两个DNA直方图参数(倍体程度和增殖指数)的解读,对哪些患者会出现肿瘤复发给出了极佳的预后判断,随访结果证实了这一点。