Zhong Xiaojian, Long Erping, Chen Wan, Xiang Wu, Liu Zhaochuan, Chen Hui, Chen Jingjing, Lin Zhuoling, Lin Haotian, Chen Weirong
State Key Laboratory of Ophthalmology, Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, Guangdong, 510060, China.
BMC Ophthalmol. 2016 Jul 8;16:100. doi: 10.1186/s12886-016-0283-4.
To compare the in-the-bag stability and visual function of single-piece intraocular lenses (IOLs) and three-piece IOLs.
A total of 65 patients with age-related cataracts (80 eyes) were enrolled and randomly assigned to receive in-the-bag implantation of either a single-piece IOL (40 eyes) or a three-piece IOL (40 eyes). Follow-up visits were conducted at 1 week, 1 month and 3 months postoperatively. Visual acuity, refraction and total aberration were examined. IOL position stability (including axial movement, decentration and tilt) was measured using a Scheimpflug imaging system.
At the 3-month follow-up visit, single-piece IOLs did not exhibit significant axial movement (0.07 ± 0.30 mm, p = 0.13) compared with their axial position at 1 week postoperatively, whereas three-piece IOLs displayed forward axial movement of -0.22 ± 0.23 mm (p < 0.0001). The mean manifest spherical equivalence (SE) of eyes with single-piece IOL was 0.15 ± 0.18D, whereas in eyes with three-piece IOLs, the mean manifest SE was -0.34 ± 0.15D (p < 0.001). There was no statistically significant difference in IOL decentration, tilt, uncorrected visual acuity, best-corrected visual acuity or total spherical aberration between the two groups.
Three months after implantation, single-piece IOLs exhibit better axial stability and more stable refractive outcome than three-piece IOLs, but both IOLs perform equally well in terms of decentration, tilt, visual acuity and total aberration.
ClinicalTrial.gov, NCT02609997 , 11/18/2015, retrospectively registered.
比较单片式人工晶状体(IOL)和三件式IOL在囊袋内的稳定性及视觉功能。
共纳入65例年龄相关性白内障患者(80只眼),随机分配接受单片式IOL(40只眼)或三件式IOL(40只眼)的囊袋内植入。术后1周、1个月和3个月进行随访。检查视力、屈光和总像差。使用Scheimpflug成像系统测量IOL位置稳定性(包括轴向移动、偏心和倾斜)。
在3个月随访时,与术后1周时的轴向位置相比,单片式IOL未表现出明显的轴向移动(0.07±0.30mm,p=0.13),而三件式IOL表现出-0.22±0.23mm的轴向向前移动(p<0.0001)。植入单片式IOL眼的平均明显球镜等效度(SE)为0.15±0.18D,而植入三件式IOL眼的平均明显SE为-0.34±0.15D(p<0.001)。两组之间IOL偏心、倾斜、未矫正视力、最佳矫正视力或总球差无统计学显著差异。
植入后3个月,单片式IOL比三件式IOL表现出更好的轴向稳定性和更稳定的屈光结果,但两种IOL在偏心、倾斜、视力和总像差方面表现相当。
ClinicalTrial.gov,NCT02609997,2015年11月18日,回顾性注册。