The clinical application of forearm and hand blood flow measurements using a gamma camera.

作者信息

Parkin A, Wilkinson D, Smye S W, Robinson P J, Garside F, Kester R C

机构信息

Department of Medical Physics, St James's University Hospital, Leeds, England.

出版信息

Surgery. 1989 Jul;106(1):26-32.

PMID:2740987
Abstract

A new method of measuring blood flow to forearms and hands during reactive hyperemia is described. Subjects are positioned with forearms and hands over a gamma camera, and the circulation below the elbows is isolated for 4 minutes with pneumatic cuffs. The remaining blood pool is labeled with technetium-99m-labeled human serum albumin given via a vein in the dorsum of the foot. The rate of rise of activity in the arms after cuff release is measured from the gradient of time activity curves, and these are calibrated for flow by counting a sample of venous blood. Blood flow per unit volume is obtained from measurements of the limb volume, done by water displacement. Results for eight normal control subjects showed a mean flow per unit volume to forearm plus hand of 32.9 +/- 6.4 ml/100 ml/min and to hand alone 36.5 +/- 6.9 ml/100 ml/min. Case studies of patients with a variety of large- and small-vessel diseases are presented, illustrating the use of the technique to aid diagnosis and measure response to treatment.

摘要

本文描述了一种测量反应性充血期间前臂和手部血流量的新方法。受试者将前臂和手置于γ相机上方,并用气动袖带将肘部以下的循环隔离4分钟。通过足背静脉注射用99m锝标记的人血清白蛋白标记剩余的血池。通过时间-活性曲线的斜率测量袖带松开后手臂中活性的上升速率,并通过对静脉血样本计数来校准血流量。单位体积的血流量通过水置换法测量肢体体积来获得。八名正常对照受试者的结果显示,前臂加手部的单位体积平均血流量为32.9±6.4ml/100ml/min,仅手部为36.5±6.9ml/100ml/min。本文还介绍了患有各种大血管和小血管疾病患者的病例研究,说明了该技术在辅助诊断和测量治疗反应方面的应用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验