Infection Control and Epidemiology Unit, Edouard Herriot Hospital, Hospices Civils de Lyon, Lyon, France.
Euro Surveill. 2016 Jul 14;21(28). doi: 10.2807/1560-7917.ES.2016.21.28.30286.
In April 2014, pulmonary Pseudomonas aeruginosa and Stenotrophomonas maltophilia co-infections potentially related to bronchoscopic procedures were identified in the intensive care units of a university hospital in Lyon, France. A retrospective cohort of 157 patients exposed to bronchoscopes from 1 December 2013 to 17 June 2014 was analysed. Environmental samples of suspected endoscopes were cultured. Bronchoscope disinfection was reviewed. Ten cases of pulmonary P. aeruginosa/S. maltophilia co-infections were identified, including two patients with secondary pneumonia. Eight cases were linked to bronchoscope A1 and two to bronchoscope A2. Cultures deriving from suction valves were positive for P. aeruginosa/S. maltophilia. Exposure to bronchoscopes A1 and A2 was independently coupled with increased risk of co-infection (adjusted odds ratio (aOR) = 84.6; 95% confidence interval (CI): 9.3-771.6 and aOR = 11.8, 95% CI: 1.2-121.3). Isolates from suction valves and clinical samples presented identical pulsotypes. The audit detected deficiencies in endoscope disinfection. No further cases occurred after discontinuation of the implicated bronchoscopes and change in cleaning procedures. This outbreak of pulmonary P. aeruginosa/S. maltophilia co-infections was caused by suction valve contamination of two bronchoscopes of the same manufacturer. Our findings underscore the need to test suction valves, in addition to bronchoscope channels, for routine detection of bacteria.
2014 年 4 月,法国里昂一所大学医院的重症监护病房发现了与支气管镜检查相关的铜绿假单胞菌和嗜麦芽窄食单胞菌合并感染。对 2013 年 12 月 1 日至 2014 年 6 月 17 日期间暴露于支气管镜的 157 名患者进行了回顾性队列分析。对疑似内镜的环境样本进行了培养。审查了支气管镜的消毒情况。共发现 10 例肺部铜绿假单胞菌/嗜麦芽窄食单胞菌合并感染病例,其中 2 例为继发性肺炎。8 例与支气管镜 A1 有关,2 例与支气管镜 A2 有关。从抽吸阀获得的培养物对铜绿假单胞菌/嗜麦芽窄食单胞菌呈阳性。使用支气管镜 A1 和 A2 与合并感染的风险增加独立相关(调整后比值比 (aOR) = 84.6;95%置信区间 (CI):9.3-771.6 和 aOR = 11.8,95% CI:1.2-121.3)。来自抽吸阀和临床样本的分离物呈现出相同的脉冲型。审核发现内镜消毒存在缺陷。停用受影响的支气管镜并改变清洗程序后,未再发生其他病例。此次铜绿假单胞菌/嗜麦芽窄食单胞菌合并感染的暴发是由同一制造商生产的两个支气管镜的抽吸阀污染引起的。我们的研究结果强调了需要测试抽吸阀,除了支气管镜通道外,还需要对细菌进行常规检测。