Lee Ki-Man, Lee Ui-Seob, Kim Eun-Hee
Department of Nuclear Engineering, Seoul National University, 1 Gwanak-ro, Gwanak-gu, Seoul, Korea.
Department of Nuclear Engineering, Seoul National University, 1 Gwanak-ro, Gwanak-gu, Seoul, Korea.
Appl Radiat Isot. 2016 Sep;115:304-311. doi: 10.1016/j.apradiso.2016.06.023. Epub 2016 Jun 21.
An alpha particle irradiator has been built in the Radiation Bioengineering Laboratory at Seoul National University (SNU) to investigate the cellular responses to alpha emissions from radon and the progeny. This irradiator is designed to have the energy of alpha particles entering target cells similar to that of alpha emissions from the radon progeny Po-218 and Po-214 residing in the human respiratory tract. For the SNU alpha particle irradiator, an irradiation system is equipped with cell dishes of 4µm thick Mylar bottom and a special setup of cells on slide for gamma-H2AX assay. Dose calibration for the alpha particle irradiator was performed by dual approaches, detection and computer simulation, in consideration of the source-to-target distance (STD) and the size of a cell dish. The uniformity of dose among cells in a dish is achieved by keeping the STD and the size of cell dish in certain ranges. The performance of the SNU alpha particle irradiator has been proven to be reliable through the gamma-H2AX assay with the human lung epithelial cells irradiated.
首尔国立大学(SNU)辐射生物工程实验室建造了一台α粒子辐照器,用于研究细胞对氡及其子体α辐射的反应。该辐照器的设计目的是使进入靶细胞的α粒子能量与存在于人类呼吸道中的氡子体Po-218和Po-214的α辐射能量相似。对于SNUα粒子辐照器,辐照系统配备了底部为4µm厚聚酯薄膜的细胞培养皿以及用于γ-H2AX检测的载玻片上细胞的特殊设置。考虑到源到靶的距离(STD)和细胞培养皿的尺寸,通过检测和计算机模拟这两种方法对α粒子辐照器进行剂量校准。通过将STD和细胞培养皿的尺寸保持在一定范围内,可实现培养皿中细胞间剂量的均匀性。通过对人肺上皮细胞进行γ-H2AX检测,已证明SNUα粒子辐照器的性能可靠。