商用牙科种植体的形貌表征

Topographic characterisation of dental implants for commercial use.

作者信息

Mendoza-Arnau A, Vallecillo-Capilla M-F, Cabrerizo-Vílchez M-Á, Rosales-Leal J-I

机构信息

University of Granada, School of Dentistry, Campus de Cartuja s/n, 18071-Granada, Spain,

出版信息

Med Oral Patol Oral Cir Bucal. 2016 Sep 1;21(5):e631-6. doi: 10.4317/medoral.20333.

Abstract

BACKGROUND

To characterize the surface topography of several dental implants for commercial use.

MATERIAL AND METHODS

Dental implants analyzed were Certain (Biomet 3i), Tissue Level (Straumann), Interna (BTI), MG-InHex (MozoGrau), SPI (Alphabio) and Hikelt (Bioner). Surface topography was ascertained using a confocal microscope with white light. Roughness parameters obtained were: Ra, Rq, Rv, Rp, Rt, Rsk and Rku. The results were analysed using single-factor ANOVA and Student-Neuman-Keuls (p<0.05) tests.

RESULTS

Certain and Hikelt obtained the highest Ra and Rq scores, followed by Tissue Level. Interna and SPI obtained lower scores, and MG-InHex obtained the lowest score. Rv scores followed the same trend. Certain obtained the highest Rp score, followed by SPI and Hikelt, then Interna and Tissue Level. MG-InHex obtained the lowest scores. Certain obtained the highest Rt score, followed by Interna and Hikelt, then SPI and Tissue Level. The lowest scores were for MG-InHex. Rsk was negative (punctured surface) in the MG-InHex, SPI and Tissue Level systems, and positive (pointed surface) in the other systems. Rku was higher than 3 (Leptokurtic) in Tissue Level, Interna, MG-InHex and SPI, and lower than 3 (Platykurtic) in Certain and Hikelt.

CONCLUSIONS

The type of implant determines surface topography, and there are differences in the roughness parameters of the various makes of implants for clinical use.

摘要

背景

对几种商用牙科种植体的表面形貌进行表征。

材料与方法

分析的牙科种植体有Certain(Biomet 3i)、组织水平型(Straumann)、Interna(BTI)、MG-InHex(MozoGrau)、SPI(Alphabio)和Hikelt(Bioner)。使用白光共聚焦显微镜确定表面形貌。获得的粗糙度参数有:Ra、Rq、Rv、Rp、Rt、Rsk和Rku。结果采用单因素方差分析和Student-Neuman-Keuls检验(p<0.05)进行分析。

结果

Certain和Hikelt的Ra和Rq得分最高,其次是组织水平型。Interna和SPI得分较低,MG-InHex得分最低。Rv得分呈现相同趋势。Certain的Rp得分最高,其次是SPI和Hikelt,然后是Interna和组织水平型。MG-InHex得分最低。Certain的Rt得分最高,其次是Interna和Hikelt,然后是SPI和组织水平型。最低得分是MG-InHex。MG-InHex、SPI和组织水平型系统的Rsk为负(表面凹陷),其他系统为正(表面尖锐)。组织水平型、Interna、MG-InHex和SPI的Rku高于3(尖顶峰),Certain和Hikelt的Rku低于3(平顶峰)。

结论

种植体类型决定表面形貌,临床使用的各种品牌种植体的粗糙度参数存在差异。