O'Brien D J, Roberts D A, Ibbott G S, Sawakuchi G O
Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030.
Elekta Limited, Crawley, West Sussex RH10 9RR, United Kingdom.
Med Phys. 2016 Aug;43(8):4915. doi: 10.1118/1.4959785.
Magnetic resonance imaging-guided radiotherapy (MRIgRT) provides superior soft-tissue contrast and real-time imaging compared with standard image-guided RT, which uses x-ray based imaging. Several groups are developing integrated MRIgRT machines. Reference dosimetry with these new machines requires accounting for the effects of the magnetic field on the response of the ionization chambers used for dose calibration. Here, the authors propose a formalism for reference dosimetry with integrated MRIgRT devices. The authors also examined the suitability of the TPR10 (20) and %dd(10)x beam quality specifiers in the presence of magnetic fields and calculated detector correction factors to account for the effects of the magnetic field for a range of detectors.
The authors used full-head and point-source Monte Carlo models of an MR-linac along with detailed detector models of an Exradin A19, an NE2571, and several PTW Farmer chambers to calculate magnetic field correction factors for six commercial ionization chambers in three chamber configurations. Calculations of ionization chamber response (performed with geant4) were validated with specialized Fano cavity tests. %dd(10)x values, TPR10 (20) values, and Spencer-Attix water-to-air restricted stopping power ratios were also calculated. The results were further validated against measurements made with a preclinical functioning MR-linac.
The TPR10 (20) was found to be insensitive to the presence of the magnetic field, whereas the relative change in %dd(10)x was 2.4% when a transverse 1.5 T field was applied. The parameters chosen for the ionization chamber calculations passed the Fano cavity test to within ∼0.1%. Magnetic field correction factors varied in magnitude with detector orientation with the smallest corrections found when the chamber was parallel to the magnetic field.
Reference dosimetry can be performed with integrated MRIgRT devices by using magnetic field correction factors, but care must be taken with the choice of beam quality specifier and chamber orientation. The uncertainties achievable under this formalism should be similar to those of conventional formalisms, although this must be further quantified.
与使用基于X射线成像的标准图像引导放疗相比,磁共振成像引导放疗(MRIgRT)具有更好的软组织对比度和实时成像能力。多个研究团队正在研发集成式MRIgRT设备。使用这些新设备进行参考剂量测定时,需要考虑磁场对用于剂量校准的电离室响应的影响。在此,作者提出了一种用于集成式MRIgRT设备参考剂量测定的形式体系。作者还研究了在存在磁场的情况下TPR10(20)和%dd(10)x射束质量规范的适用性,并计算了一系列探测器的探测器校正因子,以考虑磁场的影响。
作者使用了一个MR直线加速器的全头和点源蒙特卡罗模型,以及一个Exradin A19、一个NE2571和几个PTW Farmer电离室的详细探测器模型,来计算三种电离室配置下六个商用电离室的磁场校正因子。通过专门的法诺腔测试验证了电离室响应的计算结果(使用geant4进行)。还计算了%dd(10)x值、TPR10(20)值以及斯宾塞 - 阿蒂克斯水对空气的受限阻止本领比。结果通过与临床前运行的MR直线加速器所做测量进行进一步验证。
发现TPR10(20)对磁场的存在不敏感,而当施加横向1.5 T磁场时,%dd(10)x的相对变化为2.4%。为电离室计算所选的参数通过了法诺腔测试,误差在约0.1%以内。磁场校正因子的大小随探测器方向而变化,当电离室与磁场平行时校正最小。
通过使用磁场校正因子,可以对集成式MRIgRT设备进行参考剂量测定,但必须谨慎选择射束质量规范和电离室方向。尽管这必须进一步量化,但在这种形式体系下可实现的不确定度应与传统形式体系相似。