Suppr超能文献

[RM等弹性髋关节假体植入的5至7年随访结果]

[Results of 5-7 year follow-up of implantation of RM isoelastic hip endoprosthesis].

作者信息

Träger D

出版信息

Unfallchirurg. 1989 Jun;92(6):301-4.

PMID:2749271
Abstract

From June 1976 to June 1981, 71 isoelastic total hip endoprostheses were implanted. In a 5-to 7-year follow-up study, all but one of the hips that had been operated upon could be evaluated. In 10 cases reoperation was performed because of loosening of the stem (3 infections, 7 aseptic loosenings). In the remaining 51 endoprostheses in 45 patients, 2 were described as in poor shape. In 3 cases, an X-ray film demonstrated potential loosening of the stem. Because there is a 9.8% loosening rate of the stem, the indication for implantation has been restricted. We had no problems at all with the RM acetabular components.

摘要

1976年6月至1981年6月期间,共植入了71个等弹性全髋关节假体。在一项为期5至7年的随访研究中,除1例接受手术的髋关节外,其余均可进行评估。10例因柄松动进行了再次手术(3例感染,7例无菌性松动)。在45例患者的其余51个假体中,2个被描述为状态不佳。3例X线片显示柄有潜在松动。由于柄的松动率为9.8%,植入指征已受到限制。我们在RM髋臼组件方面没有任何问题。