Biomechanical Effects of Different Varus and Valgus Alignments in Medial Unicompartmental Knee Arthroplasty.

作者信息

Innocenti Bernardo, Pianigiani Silvia, Ramundo Gaetano, Thienpont Emmanuel

机构信息

BEAMS Department, Université Libre de Bruxelles, Brussels, Belgium.

BEAMS Department, Université Libre de Bruxelles, Brussels, Belgium; IRCCS, Istituto Ortopedico Galeazzi, Milan, Italy.

出版信息

J Arthroplasty. 2016 Dec;31(12):2685-2691. doi: 10.1016/j.arth.2016.07.006. Epub 2016 Jul 15.

Abstract

BACKGROUND

Medial unicompartmental tibial components are not always positioned following neutral mechanical alignment and a tibial varus alignment of 3° has been suggested based on several clinical follow-up studies. However, no biomechanical justification is currently available to confirm the suitability of different alignment positions.

METHODS

This study aims at quantifying the effects on bone stresses, load distribution, ligament strains, and polyethylene insert stress distribution induced by a possible varus/valgus alignment in medial unicompartmental knee arthroplasty, ranging from 6° of varus to 6° of valgus, developing and using a validated patient-specific finite element model.

RESULTS

Results demonstrate that both neutral mechanical and 3° of varus alignment induce lower stress distributions than valgus or a higher varus alignment for which higher values, up to 40%, are achieved for the polyethylene stress. When a unicompartmental knee arthroplasty is implanted, a mismatch in the stiffness of the joint is introduced, changing the load distribution from medial to lateral for all configurations with respect to the native configuration. However, slight differences are noticeable among the different configurations with a maximum of 190 N and 90 N for the lateral and the medial side, respectively.

CONCLUSION

Neutral mechanical or 3° of varus alignment present similar biomechanical outputs in the bone, collateral ligament strain, and on the polyethylene insert. A 6° varus alignment or changes in valgus alignment were always associated with more detrimental effects.

摘要

背景

内侧单髁胫骨假体并不总是按照中立机械对线进行放置,基于多项临床随访研究,有人提出胫骨内翻3°的对线方式。然而,目前尚无生物力学依据来证实不同对线位置的适用性。

方法

本研究旨在通过建立并使用经过验证的患者特异性有限元模型,量化内侧单髁膝关节置换术中可能的内翻/外翻对线(范围从内翻6°到外翻6°)对骨应力、载荷分布、韧带应变和聚乙烯衬垫应力分布的影响。

结果

结果表明,中立机械对线和3°内翻对线所产生的应力分布均低于外翻或更大程度的内翻对线,对于后者,聚乙烯应力可高达40%。当植入单髁膝关节置换假体时,会引入关节刚度不匹配的情况,相对于原始状态,所有构型的载荷分布都会从内侧转移到外侧。然而,不同构型之间存在明显差异,外侧和内侧的最大差异分别为190 N和90 N。

结论

中立机械对线或3°内翻对线在骨、侧副韧带应变以及聚乙烯衬垫上呈现相似的生物力学结果。6°内翻对线或外翻对线的改变总是伴随着更有害的影响。