Nishii Nobuhiro
Department of Cardiovascular Therapeutics, Okayama University Graduate School of Medicine, Dentistry, and Pharmaceutical Sciences, 2-5-1 Shikata-cho, Kita-ku, Okayama 700-8558, Japan.
J Arrhythm. 2016 Aug;32(4):287-92. doi: 10.1016/j.joa.2015.09.004. Epub 2015 Oct 24.
Arrhythmic management is needed after removal of cardiac implantable electronic devices (CIEDs). Patients completely dependent on CIEDs need temporary device back-up until new CIEDs are implanted. Various methods are available for device back-up, and the appropriate management varies among patients. The duration from CIED removal to implantation of a new CIED also differs among patients. Temporary pacing is needed for patients with bradycardia, a wearable cardioverter defibrillator (WCD) or catheter ablation is needed for patients with tachyarrhythmia, and sequential pacing is needed for patients dependent on cardiac resynchronization therapy. The present review focuses on arrhythmic management after CIED removal.
在移除心脏植入式电子设备(CIED)后需要进行心律失常管理。完全依赖CIED的患者在植入新的CIED之前需要临时设备支持。有多种设备支持方法可供选择,且不同患者的合适管理方式有所不同。从CIED移除到植入新CIED的时间在不同患者中也存在差异。心动过缓患者需要临时起搏,快速心律失常患者需要可穿戴式心脏复律除颤器(WCD)或导管消融,而依赖心脏再同步治疗的患者则需要顺序起搏。本综述重点关注CIED移除后的心律失常管理。