Computer Assisted Clinical Medicine, University Medical Center Mannheim, Heidelberg University, Mannheim, Germany.
Electrical and Computer Engineering, University of Minnesota, Minneapolis, Minnesota, United States.
Magn Reson Med. 2017 Aug;78(2):484-493. doi: 10.1002/mrm.26378. Epub 2016 Sep 16.
To study the feasibility of black-blood contrast in native T mapping for reduction of partial voluming at the blood-myocardium interface.
A saturation pulse prepared heart-rate-independent inversion recovery (SAPPHIRE) T mapping sequence was combined with motion-sensitized driven-equilibrium (MSDE) blood suppression for black-blood T mapping at 3 Tesla. Phantom scans were performed to assess the T time accuracy. In vivo black-blood and conventional SAPPHIRE T mapping was performed in eight healthy subjects and analyzed for T times, precision, and inter- and intraobserver variability. Furthermore, manually drawn regions of interest (ROIs) in all T maps were dilated and eroded to analyze the dependence of septal T times on the ROI thickness.
Phantom results and in vivo myocardial T times show comparable accuracy with black-blood compared to conventional SAPPHIRE (in vivo: black-blood: 1562 ± 56 ms vs. conventional: 1583 ± 58 ms, P = 0.20); Using black-blood SAPPHIRE precision was significantly lower (standard deviation: 133.9 ± 24.6 ms vs. 63.1 ± 6.4 ms, P < .0001), and blood T time measurement was not possible. Significantly increased interobserver interclass correlation coefficient (ICC) (0.996 vs. 0.967, P = 0.011) and similar intraobserver ICC (0.979 vs. 0.939, P = 0.11) was obtained with the black-blood sequence. Conventional SAPPHIRE showed strong dependence on the ROI thickness (R = 0.99). No such trend was observed using the black-blood approach (R = 0.29).
Black-blood SAPPHIRE successfully eliminates partial voluming at the blood pool in native myocardial T mapping while providing accurate T times, albeit at a reduced precision. Magn Reson Med 78:484-493, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
研究在原生 T 映射中使用黑血对比以减少血心肌界面部分容积的可行性。
在 3T 下,使用饱和脉冲准备的心脏率无关反转恢复(SAPPHIRE)T 映射序列结合运动敏感驱动平衡(MSDE)血液抑制进行黑血 T 映射。进行了体模扫描以评估 T 时间准确性。对 8 名健康受试者进行了黑血和常规 SAPPHIRE T 映射,并分析了 T 时间、精密度以及观察者内和观察者间的可变性。此外,手动绘制所有 T 图的感兴趣区(ROI),并对 ROI 厚度进行膨胀和腐蚀,以分析间隔 T 时间对 ROI 厚度的依赖性。
与常规 SAPPHIRE 相比,体模结果和体内心肌 T 时间显示出与黑血相当的准确性(体内:黑血:1562±56ms 与常规:1583±58ms,P=0.20);使用黑血 SAPPHIRE 时,精度显著降低(标准差:133.9±24.6ms 与 63.1±6.4ms,P<0.0001),且无法测量血液 T 时间。黑血序列获得了显著提高的观察者间组内相关系数(ICC)(0.996 与 0.967,P=0.011)和相似的观察者内 ICC(0.979 与 0.939,P=0.11)。常规 SAPPHIRE 显示出与 ROI 厚度的强烈依赖性(R=0.99)。使用黑血方法时,没有观察到这种趋势(R=0.29)。
黑血 SAPPHIRE 在原生心肌 T 映射中成功消除了血池的部分容积,同时提供了准确的 T 时间,尽管精密度降低。磁共振医学 78:484-493,2017。©2016 国际磁共振学会。