儿童经眶超声测量视神经鞘直径(ONSD)与有创测量颅内压(ICP)之间的关系:第一部分:重复性、观察者变异性及综合分析。

The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children : Part I: repeatability, observer variability and general analysis.

作者信息

Padayachy Llewellyn C, Padayachy Vaishali, Galal Ushma, Gray Rebecca, Fieggen A Graham

机构信息

Paediatric Neurosurgery Unit, Division of Neurosurgery, Faculty of Health Sciences, University of Cape Town, Red Cross War Memorial Children's Hospital, Cape Town, South Africa.

Trauma Unit, Red Cross War Memorial Children's Hospital, University of Cape Town, Cape Town, South Africa.

出版信息

Childs Nerv Syst. 2016 Oct;32(10):1769-78. doi: 10.1007/s00381-016-3067-5. Epub 2016 Sep 20.

Abstract

PURPOSE

The aim of this study was to investigate the relationship between optic nerve sheath diameter (ONSD) measurement and invasively measured intracranial pressure (ICP) in children.

METHODS

ONSD measurement was performed prior to invasive measurement of ICP. The mean binocular ONSD measurement was compared to the ICP reading. Physiological variables including systolic blood pressure (SBP), diastolic blood pressure (DBP), mean arterial pressure (MAP), pulse rate, temperature, respiratory rate and end tidal carbon dioxide (ETCO2) level were recorded at the time of ONSD measurement. Diagnostic accuracy analysis was performed at various ICP thresholds and  repeatability, intra- and inter-observer variability, correlation between measurements in different imaging planes as well the relationship over the entire patient cohort were examined in part I of this study.

RESULTS

Data from 174 patients were analysed. Repeatability and intra-observer variability were excellent (α = 0.97-0.99). Testing for inter-observer variability revealed good correlation (r = 0.89, p < 0.001). Imaging in the sagittal plane demonstrated a slightly better correlation with ICP (r = 0.66, p < 0.001). The ONSD measurement with the best diagnostic accuracy for detecting an ICP ≥ 20 mmHg over the entire patient cohort was 5.5 mm, sensitivity 93.2 %, specificity 74 % and odds ratio (OR) of 39.3.

CONCLUSION

Transorbital ultrasound measurement of the OSND is a reliable and reproducible technique, demonstrating a good relationship with ICP and high diagnostic accuracy for detecting raised ICP.

摘要

目的

本研究旨在探讨儿童视神经鞘直径(ONSD)测量值与有创测量的颅内压(ICP)之间的关系。

方法

在进行有创ICP测量之前先测量ONSD。将双眼ONSD测量的平均值与ICP读数进行比较。在测量ONSD时记录生理变量,包括收缩压(SBP)、舒张压(DBP)、平均动脉压(MAP)、脉搏率、体温、呼吸频率和呼气末二氧化碳(ETCO2)水平。在本研究的第一部分中,在不同的ICP阈值下进行诊断准确性分析,并检查重复性、观察者内和观察者间的变异性、不同成像平面测量值之间的相关性以及整个患者队列中的关系。

结果

分析了174例患者的数据。重复性和观察者内变异性极佳(α = 0.97 - 0.99)。观察者间变异性测试显示出良好的相关性(r = 0.89,p < 0.001)。矢状面成像与ICP的相关性略好(r = 0.66,p < 0.001)。在整个患者队列中,检测ICP≥20 mmHg诊断准确性最佳的ONSD测量值为5.5 mm,敏感性为93.2%,特异性为74%,比值比(OR)为39.3。

结论

经眶超声测量OSND是一种可靠且可重复的技术,与ICP显示出良好的关系,对检测升高的ICP具有较高的诊断准确性。