Stull J C, Schneider P J, Erenberg A, Leff R D
College of Pharmacy, University of Iowa, Iowa City 52242.
Crit Care Med. 1989 Sep;17(9):926-8. doi: 10.1097/00003246-198909000-00018.
Accurate flow from infusion pumps should be maintained when exposed to a variety of clinical conditions. The intent of this study was to evaluate in vitro flow rate accuracy of three infusion pumps subjected to the influences of variable back-pressure, solution viscosity, and infusion rates. A factorial study design was selected to determine the influence of three flow rates (5, 10, and 20 ml/h), three back-pressures (100, 200, and 300 mm Hg), and two solution viscosities (5%, 25% dextrose in water) on flow rate accuracy from three infusion pumps (Abbott 4P, IVAC 560, and Travenol 6200) using a standard gravimetric technique. Mean +/- SD accuracy values were -9.4 +/- 6.4% (range -29.1 to -0.7), 0.5 +/- 2.2% (range -4.2 to 6.3), and -0.5 +/- 4.7% (range -8.5 to 9.9) of the desired rate for the Abbott, IVAC, and Travenol devices, respectively. Back-pressure was the only factor to influence significantly flow accuracy for the Abbott device (r = .81). All factors significantly influenced accuracy for the Travenol device (r = .55). No factor influenced accuracy for the IVAC infusion pump. Both the IVAC 560 and Travenol 6200 have acceptable flow accuracy values within the range of study factors examined. The Abbott 4P had significant decreases in flow accuracy in response to increasing back-pressure.
在各种临床条件下,输液泵应保持准确的流速。本研究的目的是评估三种输液泵在可变背压、溶液粘度和输液速率影响下的体外流速准确性。采用析因研究设计,使用标准重量法确定三种流速(5、10和20毫升/小时)、三种背压(100、200和300毫米汞柱)和两种溶液粘度(5%、25%葡萄糖水溶液)对三种输液泵(雅培4P、IVAC 560和特拉文诺6200)流速准确性的影响。雅培、IVAC和特拉文诺设备的平均±标准差准确性值分别为所需速率的-9.4±6.4%(范围-29.1至-0.7)、0.5±2.2%(范围-4.2至6.3)和-0.5±4.7%(范围-8.5至9.9)。背压是影响雅培设备流速准确性的唯一因素(r = 0.81)。所有因素均显著影响特拉文诺设备的准确性(r = 0.55)。没有因素影响IVAC输液泵的准确性。在研究的因素范围内,IVAC 560和特拉文诺6200的流速准确性值均可接受。随着背压增加,雅培4P的流速准确性显著下降。