Dietscher J E, Foulks C J, Waits M
Section of Clinical Nutrition, Scott & White Clinic, Scott & White Memorial Hospital, Temple, TX 76508.
JPEN J Parenter Enteral Nutr. 1994 Jul-Aug;18(4):359-61. doi: 10.1177/014860719401800415.
There are few data on the accuracy of pumps used to deliver enteral formulas. Enteral pump manufacturers claim that accuracy is within 10% of the selected flow rate. We evaluated 15 pumps: Flexiflo Companion (n = 5), Travenol 2100 (n = 5), and Kangaroo 324 (n = 5). Three types of formula at room temperature were infused at 10, 100, and 300 mL/h. The formulas used were Jevity, Ensure Plus, and Suplena with added Promod. The Flexiflo Companion and Kangaroo 324 pumps delivered 90% to 110% of the expected volume using the ready-to-feed formulas. The Travenol 2100 pumps averaged less than 90% of the selected flow rates for the three formulas infused. All three pumps delivered less than 90% of the expected volume using a formula with an added powder, but the Travenol 2100 delivered the highest volume of this same formula. There was no significant difference in the delivery of Jevity with the Flexiflo Companion (104% +/- 5% SD) and the Kangaroo 324 pumps (99% +/- 4% SD), but the flow rate was significantly lower with the Travenol 2100 pumps (80% +/- 6% SD). There was no significant difference in the flow rate of Ensure Plus with the Kangaroo 324 (95% +/- 7% SD) and the Travenol 2100 pumps (89% +/- 4% SD). The flow rate of Ensure Plus with the Flexiflo Companion pumps was significantly higher (107% +/- 4% SD).(ABSTRACT TRUNCATED AT 250 WORDS)
关于用于输送肠内营养配方的泵的准确性的数据很少。肠内营养泵制造商声称准确性在所选流速的10%以内。我们评估了15台泵:Flexiflo Companion(n = 5)、Travenol 2100(n = 5)和袋鼠324(n = 5)。在室温下,三种类型的配方以10、100和300 mL/h的速度输注。使用的配方是加了普洛帝的能全素、安素和益力佳。Flexiflo Companion泵和袋鼠324泵使用即食配方输送的液体量为预期量的90%至110%。对于输注的三种配方,Travenol 2100泵平均输送量低于所选流速的90%。使用添加了粉末的配方时,所有三台泵输送的液体量均低于预期量的90%,但Travenol 2100泵输送的该同一种配方的液体量最高。Flexiflo Companion泵(104%±5%标准差)和袋鼠324泵(99%±4%标准差)输送能全素时没有显著差异,但Travenol 2100泵的流速显著更低(80%±6%标准差)。袋鼠324泵(95%±7%标准差)和Travenol 2100泵(89%±4%标准差)输送安素时流速没有显著差异。Flexiflo Companion泵输送安素的流速显著更高(107%±4%标准差)。(摘要截短于250字)