Department of Laboratory Medicine, Hallym University College of Medicine, Dongtan Sacred Heart Hospital, Hwaseong, Republic of Korea.
Department of Laboratory Medicine, Ewha Womans University, School of Medicine, Seoul, Republic of Korea.
Clin Chim Acta. 2017 Jan;464:1-5. doi: 10.1016/j.cca.2016.11.004. Epub 2016 Nov 5.
Since the 99th percentile reference limit for cardiac troponin (Tn) can vary depending on the reference population, Sandoval et al. published systematic selection criteria. In this study, these systematic criteria were applied for the first time to obtain the 99th percentile reference limits for 6 Tn tests.
The reference population was selected in accordance with the systematic criteria, and reference limits were set with respect to the six types of Tn assays. The coefficient of variation (CV) at the reference limit was determined using 3-4 concentrations of frozen serum.
In total, 641 South Koreans (303 males, 338 females) were selected as the reference population. The 99th percentile reference limit of Tn in the six assays ranged from 13.4 to 34.2pg/ml. The measurable fractions among the reference population ranged from 1.3% to 80.5%. The CVs at the reference limit ranged from 5.3% to 43.0%, and three were <10%.
In this study, a reference population was selected for the first time in accordance with the systematic criteria of Sandoval et al., and the reference limit for South Koreans was established. The values obtained in this study are different from those proposed by manufacturers, which confirms the importance of having a reference population. Four out of six assays did not fulfill the criteria for high-sensitivity tests.
由于心脏肌钙蛋白(Tn)的第 99 百分位参考限值可能因参考人群而异,因此 Sandoval 等人发表了系统的选择标准。在本研究中,首次应用这些系统标准来获得 6 种 Tn 检测的第 99 百分位参考限值。
根据系统标准选择参考人群,并针对 6 种 Tn 检测方法设置参考限值。使用 3-4 种冷冻血清浓度确定参考限的变异系数(CV)。
总共选择了 641 名韩国人(303 名男性,338 名女性)作为参考人群。在这 6 种检测方法中,Tn 的第 99 百分位参考限值范围为 13.4 至 34.2pg/ml。参考人群中的可测量分数范围为 1.3%至 80.5%。参考限的 CV 范围为 5.3%至 43.0%,其中 3 个低于 10%。
在本研究中,首次根据 Sandoval 等人的系统标准选择了参考人群,并为韩国人建立了参考限值。本研究中获得的值与制造商提出的值不同,这证实了拥有参考人群的重要性。6 种检测方法中有 4 种不符合高灵敏度检测的标准。