Callary S A, Field J R, Campbell D G
Royal Adelaide Hospital, Orthopaedics and Trauma, Level 4, Bice Building, North Terrace, Adelaide, South Australia, 5000, Australia.
University of Adelaide, Adelaide 5000, South Australia, Australia.
Bone Joint J. 2016 Dec;98-B(12):1604-1610. doi: 10.1302/0301-620X.98B12.37682.
The increased in vivo resistance to wear of highly crosslinked polyethylene (HXLPE) in total hip arthroplasty (THA) has led to an increased use of larger articulations which have been shown to reduce the incidence of early dislocation. To date, there are few reports of the wear of larger articulations using second generation HXLPE liners. Our prospective cohort study measured the bedding-in and early wear of large (36 mm and 40 mm diameter) articulations involving a second generation X3 HXLPE liner and compared our findings with previous clinical and in vitro studies of the same material.
The proximal penetration of the femoral head five years post-operatively was measured for 15 patients using radiostereometric analysis (RSA).
The median proximal bedding-in within the first post-operative year was 0.022 mm (interquartile range (IQR) -0.050 to 0.091). The median proximal rate of wear between one and five years was -0.004 mm/year (IQR -0.021 to 0.022). The rates of proximal, medial, 2D or 3D wear between one and five years post-operatively of the X3 HXLPE liner did not increase with larger articulations compared with our previous study of 32 mm articulations.
Although reassuring, the use of larger articulations requires continued monitoring to determine whether the low wear observed in the short-term continues to the mid- to long-term. Cite this article: Bone Joint J 2016;98-B:1604-10.
在全髋关节置换术(THA)中,高交联聚乙烯(HXLPE)体内耐磨性的提高促使更大尺寸关节的使用增加,而更大尺寸关节已被证明可降低早期脱位的发生率。迄今为止,关于使用第二代HXLPE衬垫的更大尺寸关节磨损的报道较少。我们的前瞻性队列研究测量了使用第二代X3 HXLPE衬垫的大尺寸(直径36毫米和40毫米)关节的磨合和早期磨损情况,并将我们的研究结果与之前对相同材料的临床和体外研究进行了比较。
使用放射立体测量分析(RSA)对15例患者术后五年股骨头的近端穿透情况进行测量。
术后第一年的近端磨合中位数为0.022毫米(四分位间距(IQR)为 -0.050至0.091)。术后一年至五年的近端磨损率中位数为 -0.004毫米/年(IQR为 -0.021至0.022)。与我们之前对32毫米关节的研究相比,X3 HXLPE衬垫在术后一年至五年的近端、内侧、二维或三维磨损率并未随着关节尺寸增大而增加。
尽管令人放心,但使用更大尺寸关节仍需持续监测,以确定短期观察到的低磨损情况在中长期是否依然存在。引用本文:《骨与关节杂志》2016年;98 - B:1604 - 10。