Results of a randomized controlled trial with five-year radiostereometric analysis results of vitamin E-infused highly crosslinked versus moderately crosslinked polyethylene in reverse total hip arthroplasty.

机构信息

Division of Orthopedic Surgery, Oslo University Hospital Ullevål, Oslo, Norway.

Department of Surgery, Vestfold Hospital Trust, Tønsberg, Norway.

出版信息

Bone Joint J. 2020 Dec;102-B(12):1646-1653. doi: 10.1302/0301-620X.102B12.BJJ-2020-0721.R1.

Abstract

AIMS

Vitamin E-infused highly crosslinked polyethylene (VEPE) has been introduced into total hip arthroplasty (THA) with the aim of further improving the wear characteristics of moderately and highly crosslinked polyethylenes (ModXLPE and HXLPE). There are few studies analyzing the outcomes of vitamin E-infused components in cemented arthroplasty, though early acetabular component migration has been reported. The aim of this study was to measure five-year polyethylene wear and acetabular component stability of a cemented VEPE acetabular component compared with a ModXLPE cemented acetabular component.

METHODS

In a prospective randomized controlled trial (RCT), we assessed polyethylene wear and acetabular component stability (primary outcome) with radiostereometric analysis (RSA) in 68 patients with reverse hybrid THA at five years follow-up. Patients were randomized to either a VEPE or a ModXLPE cemented acetabular component.

RESULTS

Mean polyethylene wear in the proximal direction was 0.17 mm (SD 0.15) for the VEPE group and 0.20 mm (SD 0.09) for the ModXLPE group (p = 0.005) at five years. Annual proximal wear rates were 0.03 mm/year (VEPE) and 0.04 mm/year (ModXLPE). Total 3D wear was 0.21 mm (SD 0.26) and 0.23 mm (SD 0.10) for the VEPE and ModXLPE groups, respectively (p = 0.009). Total 3D cup translation was 0.72 mm (SD 0.70) (VEPE) and 0.50 mm (SD 0.44) (ModXLPE) (p = 0.409).

CONCLUSION

At five years, there was less polyethylene wear in the VEPE group than in the ModXLPE group. Both VEPE and ModXLPE cemented components showed low annual wear rates. Component stability was similar in the two groups and remained constant up to five years. Whether these results will equate to a lower long-term revision rate is still unknown. Cite this article: 2020;102-B(12):1646-1653.

摘要

目的

维生素 E 浸渍的高度交联聚乙烯(VEPE)已被引入全髋关节置换术(THA)中,旨在进一步改善中度和高度交联聚乙烯(ModXLPE 和 HXLPE)的磨损特性。虽然已经报道了早期髋臼部件迁移,但很少有研究分析维生素 E 浸渍成分在骨水泥固定关节置换术中的结果。本研究的目的是测量骨水泥固定的 VEPE 髋臼部件与骨水泥固定的 ModXLPE 髋臼部件相比,在五年时聚乙烯的磨损和髋臼部件的稳定性。

方法

在一项前瞻性随机对照试验(RCT)中,我们使用放射立体测量分析(RSA)在 68 例接受反向混合 THA 的患者中评估聚乙烯磨损和髋臼部件稳定性(主要结果),随访时间为五年。患者被随机分配到 VEPE 或 ModXLPE 骨水泥固定髋臼组件。

结果

VEPE 组近侧聚乙烯磨损平均值为 0.17mm(SD 0.15),ModXLPE 组为 0.20mm(SD 0.09)(p=0.005),五年后。每年近端磨损率分别为 0.03mm/年(VEPE)和 0.04mm/年(ModXLPE)。VEPE 和 ModXLPE 组的总 3D 磨损分别为 0.21mm(SD 0.26)和 0.23mm(SD 0.10)(p=0.009)。VEPE 组总 3D 杯位移为 0.72mm(SD 0.70)(VEPE)和 0.50mm(SD 0.44)(ModXLPE)(p=0.409)。

结论

在五年时,VEPE 组的聚乙烯磨损比 ModXLPE 组少。VEPE 和 ModXLPE 骨水泥固定组件的年磨损率均较低。两组的部件稳定性相似,五年内保持稳定。这些结果是否会导致较低的长期翻修率仍不得而知。