使用通道化霍特林观察者对低能和中能准直器碘-123间碘苄胍(mIBG)图像中的病变可检测性进行客观比较。

Objective comparison of lesion detectability in low and medium-energy collimator iodine-123 mIBG images using a channelized Hotelling observer.

作者信息

Gregory Rebecca A, Murray Iain, Gear Jonathan, Aldridge Matthew D, Levine Daniel, Fowkes Lucy, Waddington Wendy A, Chua Sue, Flux Glenn

机构信息

Department of Physics, the Royal Marsden NHS Foundation Trust and Institute of Cancer Research, Downs Road, Sutton SM2 5PT, UK.

出版信息

Phys Med Biol. 2017 Jan 7;62(1):17-30. doi: 10.1088/1361-6560/62/1/17. Epub 2016 Dec 14.

Abstract

Iodine-123 mIBG imaging is widely regarded as a gold standard for diagnostic studies of neuroblastoma and adult neuroendocrine cancer although the optimal collimator for tumour imaging remains undetermined. Low-energy (LE) high-resolution (HR) collimators provide superior spatial resolution. However due to septal penetration of high-energy photons these provide poorer contrast than medium-energy (ME) general-purpose (GP) collimators. LEGP collimators improve count sensitivity. The aim of this study was to objectively compare the lesion detection efficiency of each collimator to determine the optimal collimator for diagnostic imaging. The septal penetration and sensitivity of each collimator was assessed. Planar images of the patient abdomen were simulated with static scans of a Liqui-Phil anthropomorphic phantom with lesion-shaped inserts, acquired with LE and ME collimators on 3 different manufacturers' gamma camera systems (Skylight (Philips), Intevo (Siemens) and Discovery (GE)). Two-hundred normal and 200 single-lesion abnormal images were created for each collimator. A channelized Hotelling observer (CHO) was developed and validated to score the images for the likelihood of an abnormality. The areas under receiver-operator characteristic (ROC) curves, Az, created from the scores were used to quantify lesion detectability. The CHO ROC curves for the LEHR collimators were inferior to the GP curves for all cameras. The LEHR collimators resulted in statistically significantly smaller Azs (p  <  0.05), of on average 0.891  ±  0.004, than for the MEGP collimators, 0.933  ±  0.004. In conclusion, the reduced background provided by MEGP collimators improved I mIBG image lesion detectability over LEHR collimators that provided better spatial resolution.

摘要

碘-123间碘苄胍(mIBG)成像被广泛认为是神经母细胞瘤和成人神经内分泌癌诊断研究的金标准,尽管用于肿瘤成像的最佳准直器尚未确定。低能(LE)高分辨率(HR)准直器具有卓越的空间分辨率。然而,由于高能光子的隔板穿透,与中能(ME)通用(GP)准直器相比,这些准直器的对比度较差。LEGP准直器提高了计数灵敏度。本研究的目的是客观比较每个准直器的病变检测效率,以确定诊断成像的最佳准直器。评估了每个准直器的隔板穿透和灵敏度。通过使用带有病变形状插入物的Liqui-Phil人体模型进行静态扫描,模拟患者腹部的平面图像,在3种不同制造商的伽马相机系统(Skylight(飞利浦)、Intevo(西门子)和Discovery(通用电气))上使用LE和ME准直器进行采集。为每个准直器创建了200张正常图像和200张单病变异常图像。开发并验证了一种通道化霍特林观察者(CHO),用于对图像的异常可能性进行评分。由这些分数生成的受试者操作特征(ROC)曲线下面积Az用于量化病变可检测性。对于所有相机,LEHR准直器的CHO ROC曲线均低于GP曲线。与MEGP准直器(平均Az为0.933±0.004)相比,LEHR准直器的Az在统计学上显著更小(p<0.05),平均为0.891±0.004。总之,与提供更好空间分辨率的LEHR准直器相比,MEGP准直器提供的背景减少提高了碘-123 mIBG图像病变的可检测性。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/9cc0/6050520/2462354b39bf/pmbaa4cb7f01_hr.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验