J Am Soc Mass Spectrom. 2017 Jun;28(6):1048-1059. doi: 10.1007/s13361-016-1562-2. Epub 2016 Dec 20.
Forensic evidentiary backlogs are indicative of the growing need for cost-effective, high-throughput instrumental methods. One such emerging technology that shows high promise in meeting this demand while also allowing on-site forensic investigation is portable mass spectrometric (MS) instrumentation, particularly that which enables the coupling to ambient ionization techniques. While the benefits of rapid, on-site screening of contraband can be anticipated, the inherent legal implications of field-collected data necessitates that the analytical performance of technology employed be commensurate with accepted techniques. To this end, comprehensive analytical validation studies are required before broad incorporation by forensic practitioners can be considered, and are the focus of this work. Pertinent performance characteristics such as throughput, selectivity, accuracy/precision, method robustness, and ruggedness have been investigated. Reliability in the form of false positive/negative response rates is also assessed, examining the effect of variables such as user training and experience level. To provide flexibility toward broad chemical evidence analysis, a suite of rapidly-interchangeable ion sources has been developed and characterized through the analysis of common illicit chemicals and emerging threats like substituted phenethylamines. Graphical Abstract ᅟ.
法庭证据积压表明,人们越来越需要经济高效、高通量的仪器方法。便携式质谱 (MS) 仪器是一种新兴技术,具有很大的应用前景,它能够满足这一需求,同时还允许在现场进行法医调查,特别是能够与环境电离技术相结合的便携式质谱仪器。虽然可以预期快速、现场筛查违禁品的好处,但现场采集数据的固有法律含义要求所采用技术的分析性能与公认技术相匹配。为此,在法医从业人员广泛采用之前,需要进行全面的分析验证研究,这也是本工作的重点。已经研究了相关的性能特征,如通量、选择性、准确性/精密度、方法稳健性和耐用性。还评估了以假阳性/阴性响应率形式的可靠性,考察了用户培训和经验水平等变量的影响。为了实现广泛的化学证据分析的灵活性,已经开发了一套快速可更换的离子源,并通过对常见非法化学品和新兴威胁(如取代苯乙胺)的分析对其进行了表征。