瞳孔散大对MAIA微视野计的影响。

Effects of pupil dilation on MAIA microperimetry.

作者信息

Han Ruofan C, Jolly Jasleen K, Xue Kanmin, MacLaren Robert E

机构信息

Oxford Eye Hospital, John Radcliffe Hospital, Oxford, UK.

Nuffield Laboratory of Ophthalmology and Oxford Biomedical Research Centre, University of Oxford, Oxford, UK.

出版信息

Clin Exp Ophthalmol. 2017 Jul;45(5):489-495. doi: 10.1111/ceo.12907. Epub 2017 Jan 24.

Abstract

BACKGROUND

Macular Integrity Assessment microperimetry assesses macular sensitivity to projected point light sources and maps eye movements to assess fixation stability. Although microperimetry is gaining prominence as an assessment tool in clinical and research settings, there is no consensus on whether it should be performed before or after pupil dilation. No studies to date have examined the effect of pupil dilation on results. The aim of this project was to elucidate the effect of pupil dilation on microperimetry outcomes.

DESIGN

Prospective audit.

PARTICIPANTS

Twenty healthy patients from postoperative cataract clinic and 10 patients with choroideremia to simulate a disease with peripheral visual field loss.

METHODS

Subjects underwent 10-2 68-point field testing using the Macular Integrity Assessment microperimeter on each eye. Subjects then underwent randomized dilation of one eye, and the test was repeated in both eyes.

MAIN OUTCOME MEASURES

We compared changes in threshold sensitivity and fixation stability pre-pupil and post-pupil dilation. The undilated eye was analysed for any learning or fatigue effect caused by test repetition.

RESULTS

Dilation produced no significant effect on threshold sensitivity (dilation effect: -0.29 decibels, P = 0.23) or fixation stability in healthy controls or in choroideremia patients (dilation effect: +0.08log bivariate contour ellipse area, P = 0.14). There was also no significant learning effect seen in the undilated eye, with no improvement in threshold sensitivity (order of eye testing: +0.03log bivariate contour ellipse area, P = 0.71).

CONCLUSIONS

In the clinical setting, patients may be tested for 10 degree microperimetry with or without pupil dilation, as both scenarios yield consistent and interchangeable results.

摘要

背景

黄斑完整性评估微视野计可评估黄斑对投射点光源的敏感度,并绘制眼动图以评估注视稳定性。尽管微视野计在临床和研究环境中作为一种评估工具正日益受到重视,但对于应在散瞳前还是散瞳后进行该检查尚无共识。迄今为止,尚无研究探讨散瞳对检查结果的影响。本项目的目的是阐明散瞳对微视野计检查结果的影响。

设计

前瞻性审计。

参与者

20名来自白内障术后门诊的健康患者以及10名患有视网膜色素变性的患者,后者用于模拟伴有周边视野缺损的疾病。

方法

受试者使用黄斑完整性评估微视野计对每只眼睛进行10-2 68点视野测试。然后,对受试者的一只眼睛进行随机散瞳,并对双眼重复该测试。

主要观察指标

我们比较了散瞳前后阈值敏感度和注视稳定性的变化。对未散瞳的眼睛分析测试重复引起的任何学习或疲劳效应。

结果

散瞳对健康对照组或视网膜色素变性患者的阈值敏感度(散瞳效应:-0.29分贝,P = 0.23)或注视稳定性均无显著影响(散瞳效应:+0.08对数双变量轮廓椭圆面积,P = 0.14)。在未散瞳的眼睛中也未观察到显著的学习效应,阈值敏感度没有改善(眼睛测试顺序:+0.03对数双变量轮廓椭圆面积,P = 0.71)。

结论

在临床环境中,无论是否散瞳,患者均可进行10度微视野计检查,因为两种情况产生的结果一致且可互换。